Noncontact Monitoring of Blood Oxygen Saturation Using Camera and Dual-Wavelength Imaging System

Authors:

Abstract
We present a noncontact method to monitor blood oxygen saturation (SpO 2 ). The method uses a CMOS camera with a trigger control to allow recording of photoplethysmography (PPG) signals alternatively at two particular wavelengths, and determines the SpO 2 from the measured ratios of the pulsatile to the nonpulsatile components of the PPG signals at these wavelengths. The signal-to-noise ratio (SNR) of the SpO 2 value depends on the choice of the wavelengths. We found that the combination of orange (λ = 611 nm) and near infrared (λ = 880 nm) provides the best SNR for the noncontact video-based detection method. This combination is different from that used in traditional contact-based SpO 2 measurement since the PPG signal strengths and camera quantum efficiencies at these wavelengths are more amenable to SpO 2 measurement using a noncontact method. We also conducted a small pilot study to validate the noncontact method over an SpO 2 range of 83%-98%. This study results are consistent with those measured using a reference contact SpO 2 device (r = 0.936, p <; 0.001). The presented method is particularly suitable for tracking one's health and wellness at home under free-living conditions, and for those who cannot use traditional contact-based PPG devices.
Funding Information
  • National Science Foundation

This publication has 24 references indexed in Scilit: