A zirconia cylindrical TM010 cavity for permittivity measurements at 1 GHz
Authors:
- 1 September 2018
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-4
Abstract
The most accurate dielectric measurements are made by resonant cavity methods, the circular TM010 type being the simplest and most common. However, an airfilled such cavity at 1 GHz needs to be 250 mm in diameter. There is another problem as well: its limited applicability with very lossy samples, due to a too low Q value. This paper describes the development and properties of a metalized zirconia ceramic cavity for use at about 1 GHz. With its permittivity ε' = 30 its diameter becomes 40 mm instead of 150 mm for the airfilled version. Additionally and importantly, the dynamic range of the loss factor ε" is greatly expanded. The calibration procedure using numerical retromodelling is described and a measurement example of a ternary alcohol mixture is carried out. The accuracy is also estimated.Keywords
This publication has 5 references indexed in Scilit:
- Techniques for measuring the microwave dielectric properties of materialsMetrologia, 2010
- Measurements of dielectric properties of foods and associated materials2009
- Retro-modelling of a dual resonant applicator and accurate dielectric properties of liquid water from −20 °C to +100 °CMeasurement Science and Technology, 2007
- Dielectric Properties of Food at 3 GHz as Determined by a Cavity Perturbation Technique.Journal of Microwave Power, 1971
- Resonances of a Microwave Cavity Partially Filled with a PlasmaJournal of Applied Physics, 1962