Dosimetric effects near implanted vascular access ports: an examination of external photon beam dose calculations
Open Access
- 28 May 2009
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 10 (3) , 3-15
Abstract
Vascular access ports are used widely in the administering of drugs for radiation oncology patients. Their dosimetric effect on radiation therapy delivery in photon beams has not been rigorously established. In this work, the effects on external beam fields when any of a variety of vascular access ports were included in the path of a high energy beam are studied. This study specifically identifies sidescatter and backscatter consequences as well as attenuation effects. The study was divided into two parts. First, a total of 18 ports underwent extended HU range CT scanning followed by 3D computer treatment planning, where homogeneous and heterogeneous plans were created for photon beams of energy 6 MV and 18 MV using a Pencil Beam Convolution (PBC) algorithm. Dose points were analyzed at locations around each device. A total of 1,440 points were reviewed in this section of the study. A replicate of the largest vascular access port was created in the treatment planning workspace for further investigation with alternative treatment planning algorithms. Then, plans were generated identical to the above and compared to the results of dose computation between the Pencil Beam Convolution algorithm, the Analytical Anisotropic Algorithm (AAA), and the EGSnrc Monte Carlo algorithm with user code DOSRZnrc (MC). A total of 300 points were reviewed in this part of the study. It was concluded that ports with more bulky construction and those with partial metal composition create the largest changes. Similar effects were observed for similar port configurations. Considerable differences between the PBC and AAA in comparison to MC are noted and discussed. By thorough examination of planning system results, the presented vascular access ports may now be ranked according to the greatest amount of change exhibited within a treatment planning system. Effects of backscatter, lateral scatter, and attenuation are up to 5.0%, 3.4% and 16.8% for 6 MV and 7.0%, 7.7% and 7.2% for 18 MV, respectively.Keywords
Funding Information
- Bard Access Systems, Inc. (# 7019)
This publication has 20 references indexed in Scilit:
- Multislice Computed Tomographic Angiography for Preinterventional Planning of Port Placement for Intra-Arterial Hepatic Infusion ChemotherapyJournal of Computer Assisted Tomography, 2007
- Complications associated with an implantable vascular access deviceJournal of Pediatric Surgery, 2006
- Dosimetric validation of the anisotropic analytical algorithm for photon dose calculation: fundamental characterization in waterPhysics in Medicine & Biology, 2006
- A 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations*Physics in Medicine & Biology, 2005
- Monte carlo dose calculations in the treatment of a pelvis with implant and comparison with pencil-beam calculationsMedical Dosimetry, 2003
- Comparison of RTP dose distributions in heterogeneous phantoms with thebeamMonte Carlo simulation systemJournal of Applied Clinical Medical Physics, 2001
- Primary mandibular reconstruction: Analysis of 64 cases and evaluation of interface radiation dosimetry on bridging platesThe Laryngoscope, 1991
- Radiation dose perturbation in the presence of permanent vascular-access injection portsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1990
- Material thrombogenicity in central venous catheterization: a comparison between soft, antebrachial catheters of silicone elastomer and polyurethaneJournal of Parenteral and Enteral Nutrition, 1984
- Radiotherapy treatment planning for patients fitted with prosthesesBritish Journal of Radiology, 1984