Impact of renal chimney intra-aortic stent length on branch and end-stent angle in chimney endovascular aneurysm repair and endovascular aneurysm sealing configurations
Open Access
- 28 December 2021
- journal article
- research article
- Published by SAGE Publications in Vascular
- Vol. 31 (2) , 234-243
Abstract
Practice patterns and durability of parallel stent graft techniques in complex endovascular aneurysm repair (EVAR) remain poorly defined. We aimed to quantify and compare the impact of renal chimney intra-aortic stent length (IASL) on geometric deformations of renal arteries in complex EVAR. Thirty-eight nonconsecutive patients underwent EVAR utilizing parallel stent graft techniques (chimney EVAR [chEVAR], n = 28; chimney endovascular aneurysm sealing [chEVAS], n = 10) between 2010 and 2016. A total of 59 renal chimney stent grafts were used. Geometric quantification was derived from three-dimensional model-based centerline extraction. Renal chimney intra-aortic stent length (IASL) was defined as the length of chimney stent that extended from the proximal edge of the chimney stent to the ostium of the corresponding renal artery. Mean IASL for both left and right renal arteries in the cohort was 35.7 mm. Renal arteries containing chimney IASL p = .06). Left renal arteries showed significantly greater branch angle among those with IASL p = .045). Mean IASL for renal arteries in chEVAR was significantly longer compared to chEVAS (39.2 vs. 26.3 mm, p = .003). No difference was noted in overall branch angle or end-stent angle based on procedure type. ChEVAR with IASL p = .03). In contrast, chEVAS patients showed no difference in end-stent angle based on IASL thresholds, but did have significantly greater branch angle among those with IASL p = .004) and right renal arteries (134.3 vs. 111.6°, p = .02). Renal chimney stents with longer IASL appear to exhibit less renal artery deformation, suggesting a more gradual and perpendicular transition of the chimney stent across the renal ostium.Keywords
This publication has 15 references indexed in Scilit:
- Three-dimensional morphometric analysis of the renal vasculatureAmerican Journal of Physiology-Renal Physiology, 2018
- Dynamic Geometric Analysis of the Renal Arteries and Aorta following Complex Endovascular Aneurysm RepairAnnals of Vascular Surgery, 2017
- Comparative geometric analysis of renal artery anatomy before and after fenestrated or snorkel/chimney endovascular aneurysm repairJournal of Vascular Surgery, 2016
- Snorkel/Chimney Stent Morphology Predicts Renal Dysfunction after Complex Endovascular Aneurysm RepairAnnals of Vascular Surgery, 2016
- Collected World Experience About the Performance of the Snorkel/Chimney Endovascular Technique in the Treatment of Complex Aortic PathologiesAnnals of Surgery, 2015
- Geometry and respiratory-induced deformation of abdominal branch vessels and stents after complex endovascular aneurysm repairJournal of Vascular Surgery, 2015
- Respiration-induced Deformations of the Superior Mesenteric and Renal Arteries in Patients with Abdominal Aortic AneurysmsJournal of Vascular and Interventional Radiology, 2013
- Respiratory‐induced 3D deformations of the renal arteries quantified with geometric modeling during inspiration and expiration breath‐holds of magnetic resonance angiographyJournal of Magnetic Resonance Imaging, 2013
- Early experience with the snorkel technique for juxtarenal aneurysmsJournal of Vascular Surgery, 2012
- Juxtarenal aortic aneurysm repairJournal of Vascular Surgery, 2010