Comparison of Correction Techniques for the Spillin Effect in Emission Tomography
- 12 March 2020
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Radiation and Plasma Medical Sciences
- Vol. 4 (4) , 422-432
Abstract
In positron emission tomography (PET) imaging, accurate clinical assessment is often affected by the partial volume effect (PVE) leading to overestimation (spill-in) or underestimation (spill-out) of activity in various small regions. The spill-in correction, in particular, can be very challenging when the target region is close to a hot background region. Therefore, this article evaluates and compares the performance of various recently developed spill-in correction techniques, namely, background correction (BC), local projection (LP), and hybrid kernelized [hybrid kernel expectation maximization (HKEM)] methods. We used a simulated digital phantom and [ 18 F]-NaF PET data of three patients with abdominal aortic aneurysms (AAAs) acquired with Siemens Biograph mMR and mCT scanners, respectively. Region of interest (ROI) analysis was performed and the extracted SUVmean, SUVmax, and target-to-background ratio (TBR) scores were compared. Results showed substantial spill-in effects from hot regions to targeted regions, which are more prominent in small structures. The phantom experiment demonstrated the feasibility of spill-in correction with all methods. For the patient data, large differences in SUVmean SUVmax and TBRmax scores were observed between the ROIs drawn over the entire aneurysm and ROIs excluding some regions close to the bone. Overall, BC yielded the best performance in spill-in correction in both phantom and patient studies.Keywords
Funding Information
- Chief Scientist Office
- Ph.D. Scholarship Award From Schlumberger Foundation Faculty for the Future
- University Research Scholarship, University of Leeds
- Medical Research Council (11/20/03)
- British Heart Foundation (FS/14/78/31020, FS/18/31/33676, CH/09/002/26360, RG/16/10/32375, RE/18/5/34216)
- Wellcome Trust (WT103782AIA)
- Royal Society Industry Fellowship (IF170011)
- EPSRC Collaborative Computational Flagship Project (EP/P022200/1)
- EPSRC Collaborative Computational Project SyneRBI (EP/T026693/1)
This publication has 36 references indexed in Scilit:
- Resolution modeling in PET imaging: Theory, practice, benefits, and pitfallsMedical Physics, 2013
- A review of partial volume correction techniques for emission tomography and their applications in neurology, cardiology and oncologyPhysics in Medicine & Biology, 2012
- Invalidity of SUV Measurements of Lesions in Close Proximity to Hot Sources due to “Shine-Through” Effect on FDG PET-CT InterpretationRadiology Research and Practice, 2012
- STIR: software for tomographic image reconstruction release 2Physics in Medicine & Biology, 2012
- 4D XCAT phantom for multimodality imaging researchMedical Physics, 2010
- Image-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PETPhysics in Medicine & Biology, 2010
- Partial-Volume Effect in PET Tumor ImagingJournal of Nuclear Medicine, 2007
- Partial Volume Correction Strategies in PETPET Clinics, 2007
- Heterogeneity of Tensile Strength and Matrix Metalloproteinase Activity in the Wall of Abdominal Aortic AneurysmsJournal of Endovascular Therapy, 2004
- A Single Scatter Simulation Technique for Scatter Correction in 3D PET1996