A micro-scale multi-frequency reactance measurement technique to detect bacterial growth at low bio-particle concentrations
- 20 March 2006
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 6 (5) , 682-692
Abstract
The technique described enables the user to detect the presence and proliferation of bacteria through an increase in the bulk capacitance (C) of the suspension, which is proportional to the bacteria count, at practical frequencies less than 1 MHz. The geometry of the micro-capillary design employed increases the bulk resistance (R) of the medium, thus increasing its RC time. This makes the measured reactance sensitive to changes in the bulk capacitance, which is usually masked by the much larger surface capacitance. The sensitivity is further enhanced by the existence of a minimum in the value of the reactance at a frequency proportional to the inverse medium RC time. The value of this reactance minimum and the frequency at which the minimum is recorded are dependent on the bacteria count and permit the detection of an initial concentration of ∼100 CFU ml−1 of E. coli within 3 hours of incubation, in comparison with the previous reported values of about 8 hours, with an initial load of 1000 CFU ml−1.Keywords
This publication has 22 references indexed in Scilit:
- Aligning fast alternating current electroosmotic flow fields and characteristic frequencies with dielectrophoretic traps to achieve rapid bacteria detectionElectrophoresis, 2005
- Application of the lag-after-pulsed-separation (LAPS) flow meter to different protein solutionsThe Analyst, 2005
- Lag-after-pulsed-separation microfluidic flowmeter for biomacromolecular solutionsSensors and Actuators B: Chemical, 2004
- Electrochemical Characterization of Polyelectrolyte Multilayers Deposited at Liquid−Liquid InterfacesLangmuir, 2003
- Ultra rapid prototyping of microfluidic systems using liquid phase photopolymerizationLab on a Chip, 2002
- Comparative evaluation of different chromogenic/fluorogenic media for detecting Escherichia coli O157:H7 in foodInternational Journal of Food Microbiology, 2001
- Measurement of charge transfer during bacterial adhesion to an indium tin oxide surface in a parallel plate flow chamberJournal of Microbiological Methods, 1999
- Impedance microbiology: quantification of bacterial content in milk by means of capacitance growth curvesJournal of Microbiological Methods, 1999
- Electrochemical changes in media due to microbialJournal of Microbiological Methods, 1984
- Ueber das Verhalten sogenannter unpolarisirbarer Elektroden gegen WechselstromAnnalen der Physik, 1899