Abstract
Introduction: Accurate PET reconstruction in spinal cord PET/MRI is challenging due to the small size of the structure and interference from background activity. The aim of this study was to establish whether MR-guided PET reconstruction can improve the accuracy of measured uptake in the spinal cord. Methods: The hybrid kernel expectation maximisation (HKEM) algorithm was evaluated on a digital anthropomorphic phantom (XNAT), and an implementation of a modified asymmetric Bowsher's prior incorporating both PET and MR data was evaluated on clinical test cases. The methods were compared against commonly used algorithms OSEM and Q.Clear. Results: The results demonstrated that the two algorithms lead to an increase in measured [ 18 F ]FDG PET tracer uptake in the spinal cord. Comparison to ground truth indicates that the improvement is insufficient to remove the bias in this small structure. Discussion: With care taken to optimise for the desired application, novel PET image reconstruction algorithms using PET and MR data to inform iterative image updates lead to improved quantification and improved image quality compared to OSEM. Further work is needed to investigate the optimal parameters and identify strategies to reduce residual bias.Keywords
Funding Information
- Medical Research Council
- Engineering and Physical Sciences Research Council
This publication has 35 references indexed in Scilit:
- SIRF: Synergistic Image Reconstruction FrameworkComputer Physics Communications, 2020
- Physiological distribution of18F-FDG in the spinal cord: A systematic reviewThe Journal of Spinal Cord Medicine, 2019
- Intercomparison of MR‐informed PET image reconstruction methodsMedical Physics, 2019
- Multi‐modal synergistic PET and MR reconstruction using mutually weighted quadratic priorsMagnetic Resonance in Medicine, 2018
- Freeform fabrication of tissue-simulating phantom for potential use of surgical planning in conjoined twins separation surgeryScientific Reports, 2017
- Standardized uptake values for [18F] FDG in normal organ tissues: Comparison of whole-body PET/CT and PET/MRIEuropean Journal of Radiology, 2013
- Magnetic Resonance-Guided Positron Emission Tomography Image ReconstructionSeminars in Nuclear Medicine, 2013
- NIH Image to ImageJ: 25 years of image analysisNature Methods, 2012
- STIR: software for tomographic image reconstruction release 2Physics in Medicine & Biology, 2012