Investigation of the stability of graphene devices for quantum resistance metrology at direct and alternating current

Authors:

Abstract
Interlaboratory comparisons of the quantized Hall resistance (QHR) are essential to verify the international coherence of primary impedance standards. Here, we report on the investigation of the stability of p-doped graphene-based QHR devices at direct and alternating currents at CMI, KRISS, and PTB. To improve the stability of the electronic transport properties of the polymer-encapsulated devices, they were shipped in an over-pressurized transport chamber. The agreement of the quantized resistance with R K/2 at direct current was on the order of 1 nΩ Ω−1 between 3.5 and 7.5 T at a temperature of 4.2 K despite changes in the carrier density during the shipping of the devices. At alternating current, the quantized resistance was realized in a double-shielded graphene Hall device. Preliminary measurements with digital impedance bridges demonstrate the good reproducibility of the quantized resistance near the frequency of 1 kHz within 0.1 μΩ Ω−1 throughout the international delivery.
Funding Information
  • National Research Foundation of Korea (NRF-2019K1A3A1A78077479)
  • Joint Research Project (18SIB07)
  • EMPIR
  • European Union

This publication has 46 references indexed in Scilit: