Comparison of Masimo Total Hemoglobin SpHb® continuous non-invasive hemoglobin monitoring device with laboratory complete blood count measurement using venous sample: Protocol for an observational substudy of the Pregnancy Risk and Infant Surveillance and Measurement Alliance Maternal and Newborn Health (PRISMA MNH) study
Open Access
- 5 February 2024
- journal article
- research article
- Published by F1000 Research Ltd in Gates Open Research
- Vol. 7, 50
Abstract
Background The Masimo Total Hemoglobin SpHb® is a continuous and non-invasive handheld device to measure hemoglobin levels. Previous research has found that SpHb is able to accurately detect hemoglobin levels in adult patients with a similar degree of bias and standard deviation to point-of-care invasive method measurements. Generally, limited clinical evidence, lack of validation of Masimo at higher than and lower than hemoglobin threshold values, and scientific consensus supporting the use of Masimo for accurate hemoglobin testing for the diagnosis of anemia during pregnancy calls for further research. Methods and analysis The proposed prospective cohort will be nested within the ongoing Pregnancy Risk and Infant Surveillance and Measurement Alliance (PRISMA) Maternal and Newborn Health (MNH) study. Three study sites (located in Zambia, Kenya, and Pakistan) will participate and collect hemoglobin data at five time points (<20 weeks, 20 weeks, 28 weeks, 36 weeks’ gestation, and six weeks postpartum). We will measure hemoglobin using a venous blood sample via hematology auto-analyzer complete blood count (gold standard) and the non-invasive device. The primary objective is to assess agreement between Masimo total hemoglobin and complete blood count and on a continuous scale using Intraclass Correlation Coefficient and Bland-Altman Analysis. The second objective is to assess agreement between the two measures on a binary scale using Positive Percentage Agreement and Negative Percentage Agreement, Cohen’s Kappa, and McNemar Test. On an ordinal scale, agreement will be measured using Weighted Cohen’s Kappa and Harrel’s Concordance Index. Lastly, we will assess factors that might affect the accuracy of Masimo total hemoglobin using linear mixed models. Conclusions The primary aim of this study is to assess the validity of the non-invasive Masimo device compared to the gold standard method of invasive hemoglobin measurements during pregnancy and postpartum periods for the diagnosis of anemia.Keywords
Funding Information
- Bill and Melinda Gates Foundation (INV-005776, INV-016221, INV-003601, INV-041999)
This publication has 16 references indexed in Scilit:
- Iron/folic acid supplementation during pregnancy prevents neonatal and under-five mortality in Pakistan: propensity score matched sample from two Pakistan Demographic and Health SurveysGlobal Health Action, 2016
- Understanding Bland Altman analysisBiochemia Medica, 2015
- Kappa statistic for clustered matched-pair dataStatistics in Medicine, 2014
- Accuracy of Noninvasive and Invasive Point-of-Care Total Blood Hemoglobin Measurement in an Outpatient SettingPostgraduate Medicine, 2012
- Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobinJournal of Clinical Monitoring and Computing, 2012
- Non-invasive haemoglobin measurement in patients undergoing elective Caesarean sectionBritish Journal of Anaesthesia, 2011
- Analysis of clustered matched‐pair dataStatistics in Medicine, 2003
- A comparison of confidence interval methods for the intraclass correlation coefficient in cluster randomized trialsStatistics in Medicine, 2002
- Successful Management of Thrombocytopenia, Microangiopathic Anemia, and Acute Renal Failure by PlasmapheresisAmerican Journal of Kidney Diseases, 1985
- Measurement in Medicine: The Analysis of Method Comparison StudiesJournal of the Royal Statistical Society: Series D (The Statistician), 1983