Limitations of quantitative oculoplethysmography and of directional Doppler ultrasonography in cerebrovascular diagnosis: assessment of an air-filled OPG system.
Authors:
- M D GinsbergMGM D Ginsberg
- S A GreenwoodSGS A Greenwood
- H I GoldbergHGH I Goldberg
- 1 January 1981
- journal article
- research article
- Published by Wolters Kluwer Health in Stroke
- Vol. 12 (1) , 27-32
Abstract
500 consecutive patients were evaluated for extracranial disease of the internal carotid arteries by an automated, air-filled, digital oculoplethysmographic system (OPG) of the Kartchner type (Zira) and by supraorbital (SO) and supratrochlear (ST) directional Doppler ultrasonography. Cerebral arteriograms were performed in 58 patients (110 vessels), and OPG timing criteria for detecting hemodynamically significant carotid artery stenosis (60% or greater diameter reduction) were ascertained. Optimal criteria were a delay of one ocular pulse, relative to the other, of greater than 12 msec; and a delay of an ocular pulse, relative to the earlier ear (external carotid) pulse, of greater than 36 msec. These criteria correctly identified 73% of vessels with 0 to 59% stenosis and 76% of vessels with 60 to 100% stenosis. However, in 26% of the vessels, OPG was either inconclusive or inaccurate. Correct diagnosis of bilateral hemodynamically significant carotid artery stenoses was made by OPG in 6 of 9 affected patients. SO Doppler was normal in 70% of vessels with 0-59% stenosis, and abnormal in 75% of vessels with 60-100% stenosis. Corresponding percentages for ST Doppler were 95% and 44%. Abnormal Doppler responses to compression of contralateral facial branches were predictive of intracranial cross-collateralization in only 25% of patients. These results suggest that both quantitative OPG in its present form and directional Doppler studies have serious limitations as non-invasive diagnostic methods.Keywords
This publication has 12 references indexed in Scilit:
- Continuous oculoplethysmographic monitoring during carotid endarterectomyThe American Journal of Surgery, 1979
- Noninvasive diagnosis of extracranial cerebrovascular diseaseNeurology, 1979
- A perspective on noninvasive diagnosis of carotid diseaseNeurology, 1979
- Doppler cerebrovascular examination: improved results with refinements in technique.Stroke, 1977
- Oculoplethysmography: An adjunct to arteriography in the diagnosis of extracranial carotid occlusive diseaseThe American Journal of Surgery, 1976
- Doppler SonographyAmerican Journal of Ophthalmology, 1974
- Noninvasive Detection and Evaluation of Carotid Occlusive DiseaseArchives of Surgery, 1973
- The diagnosis of internal carotid artery occlusion by directional Doppler sonography of the ophthalmic arteryNeurology, 1972
- Anatomic and Hemodynamic Correlations in Carotid Artery StenosisStroke, 1970
- Auscultation for Carotid Bruits in Cerebrovascular InsufficiencyJAMA, 1969