Three-Dimensional Cardiac Strain Estimation Using Spatio–Temporal Elastic Registration of Ultrasound Images: A Feasibility Study
Authors:
- An ElenAEAn Elen
- Hon Fai ChoiHCHon Fai Choi
- Dirk LoeckxDLDirk Loeckx
- Hang GaoHGHang Gao
- Piet ClausPCPiet Claus
- Paul SuetensPSPaul Suetens
- Frederik MaesFMFrederik Maes
- Jan D'HoogeJDJan D'Hooge
- 15 August 2008
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 27 (11) , 1580-1591
Abstract
Current ultrasound methods for measuring myocardial strain are often limited to measurements in one or two dimensions. Cardiac motion and deformation however are truly 3-D. With the introduction of matrix transducer technology, 3-D ultrasound imaging of the heart has become feasible but suffers from low temporal and spatial resolution, making 3-D strain estimation challenging. In this paper, it is shown that automatic intensity-based spatio-temporal elastic registration of currently available 3-D volumetric ultrasound data sets can be used to measure the full 3-D strain tensor. The method was validated using simulated 3-D ultrasound data sets of the left ventricle (LV). Three types of data sets were simulated: a normal and symmetric LV with different heart rates, a more realistic asymmetric normal LV and an infarcted LV. The absolute error in the estimated displacement was between 0.47 plusmn0.23 and 1.00 plusmn0.59 mm, depending on heart rate and amount of background noise. The absolute error on the estimated strain was 9%-21% for the radial strain and 1%-4% for the longitudinal and circumferential strains. No large differences were found between the different types of data sets. The shape of the strain curves was estimated properly and the position of the infarcts could be identified correctly. Preliminary results on clinical data taken in vivo from three healthy volunteers and one patient with an apical aneurism confirmed these findings in a qualitative manner as the strain curves obtained with the proposed method have an amplitude and shape similar to what could be expected.Keywords
This publication has 35 references indexed in Scilit:
- Experimental Validation of a New Ultrasound Method for the Simultaneous Assessment of Radial and Longitudinal Myocardial Deformation Independent of Insonation AngleCirculation, 2005
- Multiscale Motion MappingCirculation, 2004
- Strain and strain rate imaging: a new clinical approach to quantifying regional myocardial functionJournal of the American Society of Echocardiography, 2004
- Strain rate imaging using two-dimensional speckle trackingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2001
- Nonrigid registration using free-form deformations: application to breast MR imagesIEEE Transactions on Medical Imaging, 1999
- An overlap invariant entropy measure of 3D medical image alignmentPattern Recognition, 1999
- Multimodality image registration by maximization of mutual informationIEEE Transactions on Medical Imaging, 1997
- Description of the deformation of the left ventricle by a kinematic modelJournal of Biomechanics, 1992
- Texture of B-mode echograms: 3-D simulations and experiments of the effects of diffraction and scatterer densityUltrasonic Imaging, 1985
- Computerized M-mode echocardiographic analysis of left ventricular dysfunction in cardiac amyloid.Circulation, 1982