Mammographic imaging with a small format CCD‐based digital cassette: Physical characteristics of a clinical system
- 8 August 2000
- journal article
- imaging physics
- Published by Wiley in Medical Physics
- Vol. 27 (8) , 1832-1840
Abstract
The physical characteristics of a clinical charge coupled device (CCD)‐based imager (Senovision, GE Medical Systems, Milwaukee, WI) for small‐field digital mammography have been investigated. The imager employs a MinR 2000™ (Eastman Kodak Company, Rochester, NY) scintillator coupled by a 1:1 optical fiber to a front‐illuminated CCD operating at a pixel pitch of 30 microns. Objective criteria such as modulation transfer function (MTF), noise power spectrum (NPS), detective quantum efficiency (DQE), and noise equivalent quanta (NEQ) were employed for this evaluation. The results demonstrated a limiting spatial resolution (10% MTF) of 10 cy/mm. The measured DQE of the current prototype utilizing a 28 kVp, Mo–Mo spectrum beam hardened with 4.5 cm Lucite is ∼40% at close to zero spatial frequency at an exposure of 8.2 mR, and decreases to ∼28% at a low exposure of 1.1 mR. Detector element nonuniformity and electronic gain variations were not significant after appropriate calibration and software corrections. The response of the imager was linear and did not exhibit signal saturation under tested exposure conditions.Keywords
This publication has 32 references indexed in Scilit:
- Evaluating the performance of detection algorithms in digital mammographyMedical Physics, 1999
- Analysis of clustered microcalcifications by using a single numeric classifier extracted from mammographic digital imagesAcademic Radiology, 1998
- A genetic algorithm‐based method for optimizing the performance of a computer‐aided diagnosis scheme for detection of clustered microcalcifications in mammogramsMedical Physics, 1998
- Automated seeded lesion segmentation on digital mammogramsIEEE Transactions on Medical Imaging, 1998
- Digital tomosynthesis in breast imaging.Radiology, 1997
- False-positive reduction technique for detection of masses on digital mammograms: Global and local multiresolution texture analysisMedical Physics, 1997
- Density correction of peripheral breast tissue on digital mammograms.RadioGraphics, 1996
- An improved computer-assisted diagnostic scheme using wavelet transform for detecting clustered microcalcifications in digital mammogramsAcademic Radiology, 1996
- Scanned-slot digital mammography1990
- Scanned‐projection digital mammographyMedical Physics, 1987