Nyquist ghost elimination for diffusion MRI by dual-polarity readout at low b-values
Open Access
- 21 January 2025
- journal article
- research article
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 11 (2) , 027001
Abstract
Dual-polarity readout is a simple and robust way to mitigate Nyquist ghosting in diffusion-weighted echo-planar imaging but imposes doubled scan time. We here propose how dual-polarity readout can be implemented with little or no increase in scan time by exploiting an observed b-value dependence and signal averaging. The b-value dependence was confirmed in healthy volunteers with distinct ghosting at low b-values but of negligible magnitude at b = 1000 s/mm2. The usefulness of the suggested strategy was exemplified with a scan using tensor-valued diffusion encoding for estimation of parameter maps of mean diffusivity, and anisotropic and isotropic mean kurtosis, showing that ghosting propagated into all three parameter maps unless dual-polarity readout was applied. Results thus imply that extending the use of dual-polarity readout to low non-zero b-values provides effective ghost elimination and can be used without increased scan time for any diffusion MRI scan containing signal averaging at low b-values.Keywords
Funding Information
- Gustav V Jubilee Clinic Cancer Research Foundation (2020:316)
- Lion’s Cancer Research Fund of Western Sweden (2021:11)
- the Sahlgrenska University Hospital Research Funds (SU-984679)
- Stiftelsen Assar Gabrielssons Fond (FB22-29)
- The Swedish state under the agreement between the Swedish government and the county councils, the ALF-agreement (ALFGBG-942659)
- Cancerfonden (CAN 2018/628)
This publication has 25 references indexed in Scilit:
- Elimination of residual aliasing artifact that resembles brain lesion on multi‐oblique diffusion‐weighted echo‐planar imaging with parallel imaging using virtual coil acquisitionJournal of Magnetic Resonance Imaging, 2019
- Optimization of b‐value schemes for estimation of the diffusion coefficient and the perfusion fraction with segmented intravoxel incoherent motion model fittingMagnetic Resonance in Medicine, 2019
- Robust EPI Nyquist ghost removal by incorporating phase error correction with sensitivity encoding (PEC‐SENSE)Magnetic Resonance in Medicine, 2017
- Dual‐polarity GRAPPA for simultaneous reconstruction and ghost correction of echo planar imaging dataMagnetic Resonance in Medicine, 2015
- Optimization of intra-voxel incoherent motion imaging at 3.0 Tesla for fast liver examinationJournal of Magnetic Resonance Imaging, 2014
- Correction for geometric distortion and N/2 ghosting in EPI by phase labeling for additional coordinate encoding (PLACE)Magnetic Resonance in Medicine, 2007
- Removal of EPI Nyquist ghost artifacts with two‐dimensional phase correctionMagnetic Resonance in Medicine, 2004
- Artifact reduction in EPI with phase‐encoded reference scanMagnetic Resonance in Medicine, 1996
- Image reconstruction for echo planar imaging with nonequidistant k‐space samplingMagnetic Resonance in Medicine, 1992
- Echo‐planar imaging of diffusion and perfusionMagnetic Resonance in Medicine, 1991