Calibration of electron cyclotron emission radiometer for KSTAR
- 1 October 2010
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 81 (10) , 10D916
Abstract
We developed and installed an electron cyclotron emission radiometer for taking measurements of Korea Superconducting Tokamak Advanced Research (KSTAR) plasma. In order to precisely measure the absolute value of electron temperatures, a calibration measurement of the whole radiometer system was performed, which confirmed that the radiometer has an acceptably linear output signal for changes in input temperature. It was also found that the output power level predicted by a theoretical calculation agrees with that obtained by the calibration measurement. We also showed that the system displays acceptable noise-temperature performance around 0.23 eV.Keywords
This publication has 7 references indexed in Scilit:
- Development of multichannel intermediate frequency system for electron cyclotron emission radiometer on KSTAR TokamakReview of Scientific Instruments, 2008
- Effect of Ellipticity on Thermal Transport in ECH Plasmas in LHDPlasma and Fusion Research, 2008
- Development of ECE Imaging System on LHDPlasma and Fusion Research, 2007
- Radial Structure of Alfvén Eigenmodes in the DIII-D Tokamak through Electron-Cyclotron-Emission MeasurementsPhysical Review Letters, 2006
- Electron cyclotron emission radiometer upgrade on the Joint European Torus (JET) tokamakReview of Scientific Instruments, 2004
- Electron Cyclotron Emission Diagnostics in the Large Helical DeviceJournal of Plasma and Fusion Research, 2003
- Extreme broadband multichannel ECE radiometer with “zoom” deviceReview of Scientific Instruments, 2001