Performance evaluation of new Abbott Alinity hq hematology analyzer
- 4 July 2019
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 41 (5) , 657-663
Abstract
Introduction: Abbott Alinity hq is a next‐generation automated hematology analyzer providing complete blood count (CBC) with 6‐part white blood cells (WBC) differential counts. The purpose of this study was to evaluate the performance of the analyzer to verify the diagnostic and clinical utility of the Abbott Alinity hq automated system.Methods: We evaluated specimen stability, precision, linearity, carry‐over, and method comparison to assess the performance of Alinity hq. For comparison of the Alinity hq with Sysmex XN‐9000, totally 314 samples from adult and pediatric patients including both normal and abnormal hematology profiles were analyzed in parallel. The Alinity hq was also compared with the manual differential counts for the same 314 samples.Results: At 4°C, the Alinity hq analyzer showed no significant changes in CBC and WBC differential count up to 48 hours. When stored at room temperature (18‐25°C), all parameters except the mean platelet volume (MPV) were stable up to 36 hours. The Abbott Alinity hq analyzer demonstrated excellent reproducibility and between‐batch precision for all CBC and WBC differential parameters. WBC, red blood cells (RBC), hemoglobin (HGB), and platelets showed good linearity and acceptable carry‐over. Comparison with a Sysmex XN‐9000 analyzer and manual 400‐cell differential showed excellent correlation for CBC and WBC differential count parameters (correlation coefficient = 0.815‐0.999) except for mean corpuscular hemoglobin concentration (MCHC) and basophils.Conclusion: We performed initial validation studies and confirmed performance specifications on specimen stability, precision, linearity, carry‐over, and method comparison. The Abbott Alinity hq analyzer showed good analytical performance for all standard CBC parameters.Keywords
Funding Information
- Dong-A University
This publication has 16 references indexed in Scilit:
- Verification and quality control of routine hematology analyzersInternational Journal of Laboratory Hematology, 2016
- Determining the stability of complete blood count parameters in stored blood samples using the SYSMEX XE‐5000 automated haematology analyserInternational Journal of Laboratory Hematology, 2015
- Comparison of five automated hematology analyzers in a university hospital setting: Abbott Cell-Dyn Sapphire, Beckman Coulter DxH 800, Siemens Advia 2120i, Sysmex XE-5000, and Sysmex XN-2000Clinical Chemistry and Laboratory Medicine (CCLM), 2015
- ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte countingInternational Journal of Laboratory Hematology, 2014
- Evaluation of the fully automated hematological analyzer Mindray BC 6800: comparison with Horiba ABX Pentra DX120International Journal of Laboratory Hematology, 2013
- Comparison of Automated Differential Blood Cell Counts From Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in Normal and Pathologic SamplesAmerican Journal of Clinical Pathology, 2013
- Performance evaluation of the Sysmex haematology XN modular systemJournal of Clinical Pathology, 2012
- Evaluation of the Beckman Coulter UniCel DxH 800, Beckman Coulter LH 780, and Abbott Diagnostics Cell-Dyn Sapphire Hematology Analyzers on Adult Specimens in a Tertiary Care HospitalAmerican Journal of Clinical Pathology, 2011
- Aging stability of complete blood count and white blood cell differential parameters analyzed by Abbott CELL‐DYN Sapphire hematology analyzerInternational Journal of Laboratory Hematology, 2009
- European Multi-Center Evaluation of the Abbott Cell-Dyn Sapphire Hematology AnalyzerLaboratory Hematology, 2006