Intubation in the operating theatre using the Video-Airtraq™ laryngoscope in difficult circumstances by a face-to-face tracheal intubation technique
Authors:
Open Access
- 1 June 2014
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 112 (6) , 1118-1119
Abstract
Editor—We have demonstrated that the Video-Airtraq™ laryngoscope (VAL; VYGON, Ecouen, France) used in difficult airway management simulation conditions, allowed fast and easy face-to-face oro-tracheal intubation (ftf-OTI) with the difficult airway manikin placed in sitting position facing the operator.1 After both manikin and clinical training, and after a clinical evaluation trial, we have incorporated VAL-ftf-OTI technique in our difficult airway management algorithm. We propose the VAL-ftf-OTI technique as a standard to elective/emergency operating theatre patients showing an inter-incisor distance or mouth aperture of >25 mm, necessitating a second tracheal intubation (re-intubation) after the first one, attempted with the Macintosh laryngoscope (ML), had been difficult, as defined by an intubation difficulty score2 (IDS) of >5, in the condition of a difficult airway management algorithm.3Keywords
This publication has 5 references indexed in Scilit:
- Simulating face-to-face tracheal intubation of a trapped patient: a randomized comparison of the LMA Fastrach™, the GlideScope™, and the Airtraq™ laryngoscopeBritish Journal of Anaesthesia, 2011
- An Algorithm for Difficult Airway Management, Modified for Modern Optical Devices (Airtraq Laryngoscope; LMA CTrach ™)Anesthesiology, 2011
- Tracheal intubation of morbidly obese patients: a randomized trial comparing performance of Macintosh and Airtraq™ laryngoscopesBritish Journal of Anaesthesia, 2008
- Tracheal Intubation Using the Airtraq® in Morbid Obese Patients Undergoing Emergency Cesarean DeliveryAnesthesiology, 2007
- The Intubation Difficulty Scale (IDS)Anesthesiology, 1997