Radiation shielding and safety implications following linac conversion to an electron FLASH‐RT unit
Authors:
- 12 August 2021
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 48 (9) , 5396-5405
Abstract
Purpose: Due to their finite range, electrons are typically ignored when calculating shielding requirements in megavoltage energy linear accelerator vaults. However, the assumption that 16 MeV electrons need not be considered does not hold when operated at FLASH‐RT dose rates (~200× clinical dose rate), where dose rate from bremsstrahlung photons is an order of magnitude higher than that from an 18 MV beam for which shielding was designed. We investigate the shielding and radiation protection impact of converting a Varian 21EX linac to FLASH‐RT dose rates.Methods: We performed a radiation survey in all occupied areas using a Fluke Biomedical Inovision 451P survey meter and a Wide Energy Neutron Detection Instrument (Wendi)‐2 FHT 762 neutron detector. The dose rate from activated linac components following a 1.8‐min FLASH‐RT delivery was also measured.Results: When operated at a gantry angle of 180° such as during biology experiments, the 16 MeV FLASH‐RT electrons deliver ~10 µSv/h in the controlled areas and 780 µSv/h in the uncontrolled areas, which is above the 20 µSv in any 1‐h USNRC limit. However, to exceed 20 µSv, the unit must be operated continuously for 92 s, which corresponds in this bunker and FLASH‐RT beam to a 3180 Gy workload at isocenter, which would be unfeasible to deliver within that timeframe due to experimental logistics. While beam steering and dosimetry activities can require workloads of that magnitude, during these activities, the gantry is positioned at 0° and the dose rate in the uncontrolled area becomes undetectable. Likewise, neutron activation of linac components can reach 25 µSv/h near the isocenter following FLASH‐RT delivery, but dissipates within minutes, and total doses within an hour are below 20 µSv.Conclusion: Bremsstrahlung photons created by a 16 MeV FLASH‐RT electron beam resulted in consequential dose rates in controlled and uncontrolled areas, and from activated linac components in the vault. While our linac vault shielding proved sufficient, other investigators would be prudent to confirm the adequacy of their radiation safety program, particularly if operating in vaults designed for 6 MV.Keywords
This publication has 31 references indexed in Scilit:
- Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in miceScience Translational Medicine, 2014
- Intercomparison of radiation protection instrumentation in a pulsed neutron fieldNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014
- Dose rate dependence for different dosimeters and detectors: TLD, OSL, EBT films, and diamond detectorsMedical Physics, 2012
- APENELOPE‐based system for the automated Monte Carlo simulation of clinacs and voxelized geometries—application to far‐from‐axis fieldsMedical Physics, 2011
- TENTH VALUE LAYERS FOR 60Co GAMMA RAYS AND FOR 4, 6, 10, 15, AND 18 MV X RAYS IN CONCRETE FOR BEAMS OF CONE ANGLES BETWEEN 0° AND 14° CALCULATED BY MONTE CARLO SIMULATIONHealth Physics, 2007
- Charge collection efficiency in ionization chambers exposed to electron beams with high dose per pulsePhysics in Medicine & Biology, 2006
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Primary and Leakage Radiation Calculations at 6, 10 and 25 MeVHealth Physics, 1984
- X-RAY DEPTH DOSES FROM LINEAR ACCELERATORS IN THE ENERGY RANGE FROM 10 TO 32 MeVAmerican Journal of Roentgenology, 1975
- DESIGN OF X-RAY TARGETS FOR HIGH ENERGY LINEAR ACCELERATORS IN RADIOTHERAPYAmerican Journal of Roentgenology, 1974