Virtual-Pinhole PET
Open Access
- 20 February 2008
- journal article
- Published by Society of Nuclear Medicine in Journal of Nuclear Medicine
- Vol. 49 (3) , 471-479
Abstract
We proposed and tested a novel geometry for PET system design analogous to pinhole SPECT called the virtual-pinhole PET (VP-PET) geometry to determine whether it could provide high-resolution images. Methods: We analyzed the effects of photon acolinearity and detector sizes on system resolution and extended the empiric formula for reconstructed image resolution of conventional PET proposed earlier to predict the resolutions of VP-PET. To measure the system resolution of VP-PET, we recorded coincidence events as a 22Na point source was stepped across the coincidence line of response between 2 detectors made from identical arrays of 12 × 12 lutetium oxyorthosilicate crystals (each measuring 1.51 × 1.51 × 10 mm3) separated by 565 mm. To measure reconstructed image resolution, we built 4 VP-PET systems using 4 types of detectors (width, 1.51–6.4 mm) and imaged 4 point sources of 64Cu (half-life = 12.7 h to allow a long acquisition time). Tangential and radial resolutions were measured and averaged for each source and each system. We then imaged a polystyrene plastic phantom representing a 2.5-cm-thick cross-section of isolated breast volume. The phantom was filled with an aqueous solution of 64Cu (713 kBq/mL) in which the following were imbedded: 4 spheric tumors ranging from 1.8 to 12.6 mm in inner diameter (ID), 6 micropipettes (0.7- or 1.1-mm ID filled with 64Cu at 5×, 20×, or 50× background), and a 10.0-mm outer-diameter cold lesion. Results: The shape and measured full width at half maximum of the line spread functions agree well with the predicted values. Measured reconstructed image resolution (2.40–3.24 mm) was ±6% of the predicted value for 3 of the 4 systems. In one case, the difference was 12.6%, possibly due to underestimation of the block effect from the low-resolution detector. In phantom experiments, all spheric tumors were detected. Small line sources were detected if the activity concentration is at least 20× background. Conclusion: We have developed and characterized a novel geometry for PET. A PET system following the VP-PET geometry provides high-resolution images for objects near the system's high-resolution detectors. This geometry may lead to the development of special-purpose PET systems or resolution-enhancing insert devices for conventional PET scanners.Keywords
This publication has 24 references indexed in Scilit:
- 2D linear and iterative reconstruction algorithms for a PET-insert scannerPhysics in Medicine & Biology, 2007
- Instrumentation Aspects of Animal PETAnnual Review of Biomedical Engineering, 2005
- Design and characterization of IndyPET-II:A high-resolution, high-sensitivity dedicated research scannerIEEE Transactions on Nuclear Science, 2003
- Performance evaluation of the microPET P4: a PET system dedicated to animal imagingPhysics in Medicine & Biology, 2001
- Techniques to improve the spatial sampling of MicroPET-a high resolution animal PET tomographIEEE Transactions on Nuclear Science, 2000
- MicroPET: a high resolution PET scanner for imaging small animalsIEEE Transactions on Nuclear Science, 1997
- CT fan-beam parametrizations leading to shift-invariant filteringInverse Problems, 1996
- Simple charge division readouts for imaging scintillator arrays using a multi-channel PMTIEEE Transactions on Nuclear Science, 1996
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1982
- On the Angular Distribution of Two-Photon Annihilation RadiationPhysical Review B, 1950