Abstract
We report the utilization of transparent poly(dimethylsiloxane) (PDMS) microwires as the transducer for optical gas flowmetry. The elasticity of the PDMS microwire was exploited not only to miniaturize and simplify the flowm eter but also to widen and tune the measurement range through mechanical stretching. Using a 9 mm long microwire, we achieved 2.8∼9.8 dB/SLM sensitivity. A 500 μm stretching of the microwire also shifted the measurement range from 1 to 4 SLM. The experimental results agreed well with predictions based on the fluid dynamic/optical model.
Funding Information
  • National Science Foundation (NSF) (ECCS-0954845)

This publication has 15 references indexed in Scilit: