Contamination of the medullary canal following pin‐tract infection
Authors:
- 1 November 1999
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 17 (6) , 947-952
Abstract
We developed an ovine model of an external‐fixator pin‐tract infection. With use of a novel method of tissue sampling, infection of the medullary canal was confirmed in all (10 of 10) external‐fixator pins that were contaminated with Staphylococcus aureus after they were inserted. In addition, all (five of five) adjacent, uncontaminated pins became infected. We demonstrated that pin‐tract infection can be difficult to diagnose clinically, despite gross infection of the tract, and that bacteria can spread within the medulla. Three of the infected pins (20%) did not appear clinically infected, were not loose, and were normal on radiographs after 1 week. Staphylococcus aureus was isolated from the medulla around all 15 pin tracts, and nine other organisms were isolated from the tracts. Despite the presence of infection, the majority of the pins remained well fixed in the bone after 2 weeks.Keywords
This publication has 23 references indexed in Scilit:
- Novel animal model for studying the molecular mechanisms of bacterial adhesion to bone‐implanted metallic devices: Role of fibronectin in Staphylococcus aureus adhesionJournal of Orthopaedic Research, 1996
- The prevention of infection in open fractures: an experimental study of the effect of fracture stabilityInjury, 1994
- Induction and Prevention of Pin Loosening in ExternalJournal of Orthopaedic Trauma, 1991
- An in vivo model to study the pathobiology of infectious biofilms on biomaterial surfacesJournal of Biomedical Materials Research, 1991
- Tibia Nonunions Treated by Interlocked NailingJournal of Orthopaedic Trauma, 1990
- Bacterial Colonization of Orthopedic Fixation Devices in the Absence of Clinical InfectionThe Journal of Infectious Diseases, 1988
- Pin tract infections: A canine modelJournal of Orthopaedic Research, 1987
- Heat caused by drilling cortical bone: Temperature measuredin vivoin patients and animalsActa Orthopaedica, 1984
- EXTERNAL PIN TRANSFIXION OF FRACTURESAnnals of Surgery, 1944
- THE STADER REDUCTION SPLINTAnnals of Surgery, 1942