Improving detection thresholds and robust event filtering in single-particle and single-cell ICP-MS analysis
- 30 October 2023
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 38 (12) , 2607-2618
Abstract
In this work, a modular data processing workflow for single-particle (sp) and single-cell (sc) inductively coupled plasma-mass spectrometry (ICP-MS) is presented. To achieve more reliable detection thresholds, a special focus is placed on the parameter estimation of Gaussian and Poisson distributions that describe the background (BG) signal. For Gaussian models, the widely used iterative outlier test was improved by an algorithm that adjusts the test for different BG levels by incrementing the test factor. Through careful evaluation, the standard deviation of the experimental sc- and sp-ICP-MS data was applied as a robust measure of the convergence quality of the test. In addition, the outlier analysis was separated from the subsequent event detection more strictly than it is often reported. Importantly, a data-dependent decision criterion based on Gaussian and Poisson modeling was developed to effectively address extra-Poisson variance in experimental data. In the second part, a gate filter was developed to reduce the amount of excess false-positive events in sc-ICP-MS. To that end, a secondary filter based on the signal peak height is used to remove rare false-positive events without affecting the signal intensity of the events that are detected correctly. Two approaches, based on a numerical approximation via the detection limit, and critical values of the Gaussian and Poisson distribution are presented to calculate the gate filter level. Possible sources of false-positive events, some of which are specific to sc-ICP-MS, are discussed. The combined processing workflow was applied to analyze the distribution of six endogenous elements in Chlamydomonas reinhardtii cell populations. The gate filter corrected the cell number concentration by up to 44% (22% on average), and mass per cell by up to 30% (17% on average).Keywords
Funding Information
- Deutsche Forschungsgemeinschaft (CRC 1450 – 431460824)
This publication has 45 references indexed in Scilit:
- Size Discrimination and Detection Capabilities of Single-Particle ICPMS for Environmental Analysis of Silver NanoparticlesAnalytical Chemistry, 2012
- Determining Transport Efficiency for the Purpose of Counting and Sizing Nanoparticles via Single Particle Inductively Coupled Plasma Mass SpectrometryAnalytical Chemistry, 2011
- Selective identification, characterization and determination of dissolved silver(i) and silver nanoparticles based on single particle detection by inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 2011
- Shorter signals for improved signal to noise ratio, the influence of Poisson distributionJournal of Analytical Atomic Spectrometry, 2010
- Gold colloid analysis by inductively coupled plasma-mass spectrometry in a single particle modeAnalytica Chimica Acta, 2006
- Zirconia colloid analysis by single particle inductively coupled plasma–mass spectrometryAnalytica Chimica Acta, 2004
- Masking and swamping effects on tests for multiple outliers in normal sampleCommunications in Statistics - Theory and Methods, 1989
- Percentage Points for a Generalized ESD Many-Outlier ProcedureTechnometrics, 1983
- Some Grubbs-Type Statistics for the Detection of Several OutliersTechnometrics, 1972
- Procedures for Detecting Outlying Observations in SamplesTechnometrics, 1969