Investigation of dose homogeneity in paediatric anthropomorphic phantoms for a simple total body irradiation technique.
Authors:
- B PerrinBPB Perrin
- J M WilkinsonJWJ M Wilkinson
- H R GattamaneniHGH R Gattamaneni
- D P DeakinDDD P Deakin
- 1 March 2000
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 73 (867) , 317-321
Abstract
The technique for treating total body irradiation patients used at the centre involves no compensation for the inhomogeneity of patient shape. Dose is prescribed to the lung, and monitor units are derived from standard data depending on the external dimensions of the patient at nipple level. Dose measurements were made during standard treatments on three paediatric anthropomorphic phantoms representing children of 5, 10 and 15 years of age. The results confirmed that the measured dose to the lung was within 4% of the prescribed dose, and dose homogeneity was within ¡5%, excluding the neck, where the higher measured doses were still within tissue tolerance.Keywords
This publication has 3 references indexed in Scilit:
- Total body irradiation. Report on a meeting organized by the BIR Oncology Committee, held at The Royal Institute of British Architects, London, 28 November 1996.British Journal of Radiology, 1997
- Quality assurance in radiotherapy by in vivo dosimetry. 2. Determination of the target absorbed doseRadiotherapy and Oncology, 1990
- Quality assurance in radiotherapy by in vivo dosimetry. 1. Entrance dose measurements, a reliable procedureRadiotherapy and Oncology, 1990