Comparison of conventional twist drill protocol and piezosurgery for implant insertion: an ex vivo study on different bone types
- 22 January 2016
- journal article
- research article
- Published by Wiley in Clinical Oral Implants Research
- Vol. 28 (2) , 207-213
Abstract
Objective: The aim of this ex vivo study was to compare implant insertion procedures using piezosurgery and conventional drilling in different qualities of bone. Implant bed preparation time, generated heat, and primary implant stability were analyzed.Material and methods: Fresh ex vivo porcine bone block samples (cancellous, mixed, and cortical bone) were obtained. The bone quality was quantified by ultrasound transmission velocity (UTV). Each bone sample received three implants of the same diameter using each of the techniques of piezosurgery and conventional twist drills. Time for preparation was taken and the temperature while performing the osteotomy was measured using infrared spectroscopy. The primary implant stability after osteotomy was measured using resonance frequency analysis (RFA) and extrusion torque (ET). ANOVA with post hoc Tukey test was carried out to compare the values for the three different groups.Results: The UTV values strongly correlated with the density of the bone samples. There was a significant increase in time (threefold, P < 0.05 [302 s vs. 122 s in cortical bone]) but no difference in the temperature for the piezo group (~37°C in cortical bone). Regardless of the osteotomy technique, there was a statistically significant increase in RFA and ET values in implants inserted in cancellous bone (RFA: piezo 77, drill 76; ET: piezo 22, drill 21), mixed bone (RFA: piezo 85, drill 86; ET: piezo 105, drill 61), and cortical bone (RFA: piezo 90, drill 87; ET piezo 184, drill 79) samples, respectively (P < 0.05). In between the different osteotomy groups, there was no difference in the RFA values but significant higher ET values in mixed/cortical bone samples in favor for the piezosurgery group.Conclusion: Piezosurgery and conventional implant bed drilling procedure do have similar mechanical outcomes regarding primary stability with high RFA values, but the preparation does need more time for piezosurgery group, so that piezosurgery might be a valuable tool in only very specific cases for implant bed preparation.Keywords
This publication has 24 references indexed in Scilit:
- Evaluation of ultrasound transmission velocity and 3-dimensional radiology in different bone types for dental implantology: a comparative ex vivo studyOral Surgery, Oral Medicine, Oral Pathology and Oral Radiology, 2013
- Dental implant placement after mandibular reconstruction by microvascular free fibula flap: Current knowledge and remaining questionsOral Oncology, 2011
- Piezoelectric vs. conventional implant site preparation: ex vivo implant primary stabilityClinical Oral Implants Research, 2011
- Relation between bone quality values from ultrasound transmission velocity and implant stability parameters - an ex vivo studyClinical Oral Implants Research, 2011
- Ultrasound Transmission Velocity for Noninvasive Evaluation of Jaw Bone Quality In Vivo Before Dental ImplantationUltrasound in Medicine & Biology, 2008
- Micromorphometrical analysis of conventional osteotomy techniques and ultrasonic osteotomy at the rabbit skullClinical Oral Implants Research, 2008
- Dental Implantation: Ultrasound Transmission Velocity to Evaluate Critical Bone Quality - An Animal ModelUltraschall in der Medizin - European Journal of Ultrasound, 2007
- Heat Production by 3 Implant Drill Systems After Repeated Drilling and SterilizationJournal of Oral and Maxillofacial Surgery, 2006
- Perspectives: Ultrasound assessment of boneJournal of Bone and Mineral Research, 1993
- Temperature threshold levels for heat-induced bone tissue injury: A vital-microscopic study in the rabbitThe Journal of Prosthetic Dentistry, 1983