The capsule's contribution to total hip construct stability – A finite element analysis
Authors:
Open Access
- 14 April 2011
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 29 (11) , 1642-1648
Abstract
Instability is a significant concern in total hip arthroplasty (THA), particularly when there is structural compromise of the capsule due to pre‐existing pathology or due to necessities of surgical approach. An experimentally grounded fiber‐direction‐based finite element model of the hip capsule was developed, and was integrated with an established three‐dimensional model of impingement/dislocation. Model validity was established by close similarity to results from a cadaveric experiment in a servohydraulic hip simulator. Parametric computational runs explored effects of graded levels of capsule thickness, of regional detachment from the capsule's femoral or acetabular insertions, of surgical incisions of capsule substance, and of capsule defect repairs. Depending strongly upon the specific site, localized capsule defects caused varying degrees of construct stability compromise, with several specific situations involving over 60% decrement in dislocation resistance. Construct stability was returned substantially toward intact‐capsule levels following well‐conceived repairs, although the suture sites involved were often at substantial risk of failure. These parametric model results underscore the importance of retaining or robustly repairing capsular structures in THA, in order to maximize overall construct stability. © 2011 Orthopaedic Research Society Published by Wiley Periodicals, Inc. J Orthop Res 29:1642–1648, 2011Keywords
This publication has 31 references indexed in Scilit:
- Hyperelastic modelling of arterial layers with distributed collagen fibre orientationsJournal of The Royal Society Interface, 2005
- Design factors influencing performance of constrained acetabular liners: finite element characterizationJournal of Biomechanics, 2005
- Flexor digitorum profundus tendon to bone tunnel repair: A vascularization and histologic study in caninesThe Journal of Hand Surgery, 2005
- Range of Motion and Stability in Total Hip Arthroplasty With 28-, 32-, 38-, and 44-mm Femoral Head SizesThe Journal of Arthroplasty, 2005
- Activity-dependence of the “safe zone” for impingement versus dislocation avoidanceMedical Engineering & Physics, 2004
- Failure of Capsular Enhanced Short External Rotator Repair After Total Hip ReplacementClinical Orthopaedics and Related Research, 2004
- Spatial distribution of hip capsule structural and material propertiesJournal of Biomechanics, 2002
- Development and Physical Validation of a Finite Element Model of Total Hip DislocationComputer Methods in Biomechanics and Biomedical Engineering, 1999
- Dislocation in low-friction arthroplastyThe Journal of Arthroplasty, 1992
- Computed Tomography of the Hip Joint CapsuleJournal of Computer Assisted Tomography, 1983