Agreement between extrapolated corneoscleral topographical data obtained during natural and retracted eyelid positions
Open Access
- 21 February 2025
- journal article
- Published by Springer Nature in Ophthalmic and Physiological Optics
- Vol. 45 (3) , 627-636
Abstract
Purpose: To examine the influence of aperture size (corneoscleral data coverage) on extrapolated scleral sagittal height data generated by the Pentacam HR derived from the Corneo Scleral Profile (CSP) software, by comparing measurements obtained during natural and retracted eyelid positions. Methods: Corneoscleral topography of the left eye of 20 young (age: 22 [3] years) healthy adults with normal corneas was measured using the Pentacam HR CSP with the eyelids in their natural primary gaze resting position and during eyelid retraction with a wire speculum. The measured and extrapolated sagittal height data were exported from the instrument and analysed over a range of chord diameters (10.0, 12.5, 15.0 and 17.5 mm) and locations (superior, inferior, nasal and temporal) using customised software. Results: Eyelid retraction increased data coverage (% of available data points within 360°) for the 12.5 and 15.0 mm chord diameters (by 25% and 35%, respectively, p < 0.001), but by less than 10% for the 10.0 and 17.5 mm chord diameters. Significant differences in extrapolated sagittal height data were observed between the natural and retracted eyelid positions with respect to chord diameter and location (both p < 0.001), with the greatest difference observed superiorly for a 17.5 mm chord diameter (mean extrapolated sagittal height difference, retracted minus natural: −447 ± 401 μm, p < 0.0001). Conclusions: Eyelid retraction substantially increased data coverage for the 12.5 and 15.0 mm chord diameters. Significant differences in the extrapolated sagittal height data generated from measurements obtained during natural and retracted eyelid positions were observed for the superior location (12.5, 15.0 and 17.5 mm chord diameters) and inferior and nasal locations (17.5 mm chord diameter). Extrapolated sagittal height values obtained during the natural eyelid position were typically greater than those obtained with eyelid retraction, most likely due to the amount of measured data available for extrapolation.Keywords
This publication has 27 references indexed in Scilit:
- The effect of extended periods of mini-scleral lens wear on the conjunctival/episcleral and scleral thicknessContact Lens and Anterior Eye, 2024
- Scleral lens prescription and management practices: Emerging consensusContact Lens and Anterior Eye, 2023
- Comparing sagittal heights calculated using corneal parameters and those measured with profilometryContact Lens and Anterior Eye, 2022
- Conjunctival prolapse during open eye scleral lens wearContact Lens and Anterior Eye, 2021
- Anatomical and physiological considerations in scleral lens wear: Conjunctiva and scleraContact Lens and Anterior Eye, 2020
- The influence of centre thickness on miniscleral lens flexureContact Lens and Anterior Eye, 2019
- Anterior eye surface changes following miniscleral contact lens wearContact Lens and Anterior Eye, 2019
- Corneo-scleral limbal changes following short-term soft contact lens wearContact Lens and Anterior Eye, 2017
- Surface coverage with single vs. multiple gaze surface topography to fit scleral lensesContact Lens and Anterior Eye, 2017
- Morphological changes in the conjunctiva, episclera and sclera following short-term miniscleral contact lens wear in rigid lens neophytesContact Lens and Anterior Eye, 2016