Experimental energy resolution of a paracentric hemispherical deflector analyzer for different entry positions and bias
- 1 April 2013
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 84 (4) , 043105
Abstract
A specially designed hemispherical deflector analyzer (HDA) with 5-element input lens having a movable entry position R0 suitable for electron energy analysis in atomic collisions was constructed and tested. The energy resolution of the HDA was experimentally determined for three different entry positions R0 = 84, 100, 112 mm as a function of the nominal entry potential V(R0) under pre-retardation conditions. The resolution for the (conventional) entry at the mean radius R0 = 100 mm was found to be a factor of 1.6-2 times worse than the resolution for the two (paracentric) positions R0 = 84 and 112 mm at particular values of V(R0). These results provide the first experimental verification and a proof of principle of the utility of such a paracentric HDA, while demonstrating its advantages over the conventional HDA: greater dispersion with reduced angular aberrations resulting in better energy resolution without the use of any additional fringing field correction electrodes. Supporting simulations of the entire lens plus HDA spectrometer are also provided and mostly found to be within 20%–30% of experimental values. The paracentric HDA is expected to provide a lower cost and/or more compact alternative to the conventional HDA particularly useful in modern applications utilizing a position sensitive detector.Keywords
This publication has 16 references indexed in Scilit:
- Fringing field optimization of hemispherical deflector analyzers using BEM and FDMJournal of Electron Spectroscopy and Related Phenomena, 2010
- The hemispherical deflector analyser revisitedJournal of Electron Spectroscopy and Related Phenomena, 2008
- Design, simulation and construction of an electron–electron coincidence spectrometerJournal of Electron Spectroscopy and Related Phenomena, 2007
- Novel and traditional fringing field correction schemes for the hemispherical analyser: comparison of first-order focusing and energy resolutionMeasurement Science and Technology, 2007
- Investigation of the accuracy of ion optics simulations using Kepler orbits in a spherical capacitorInternational Journal of Mass Spectrometry, 2007
- Using the fringing fields of a hemispherical spectrograph to improve its energy resolutionMeasurement Science and Technology, 2006
- Optimal energy resolution of a hemispherical analyzer with virtual entryApplied Physics Letters, 2005
- The hemispherical deflector analyser revisited. I. Motion in the ideal 1/r potential, generalized entry conditions, Kepler orbits and spectrometer basic equationJournal of Electron Spectroscopy and Related Phenomena, 2002
- Coincidence studies of the influence of resonances on simultaneous ionization-excitation of helium by electron impactJournal of Physics B: Atomic, Molecular and Optical Physics, 2002
- Simultaneous excitation-ionization of helium to the stateJournal of Physics B: Atomic, Molecular and Optical Physics, 1998