Abstract
A HPLC method was developed and validated to analyze meropenem and vaborbactam simultaneously in murine plasma and saline matrixes. A 60-μL volume of extracted sample was injected onto a 5-μm BDS Phenyl-Hypersil C18 reversed-phase column and analyzed with a UV detector set at 298 nm for the first 4.9 min and switched to 240 nm. The mobile phase contained a mixture of methanol and 25-mM sodium phosphate buffer set at a flow rate of 1.0 mL/min for the 16 min run time. Cefuroxime was used as the internal standard. The standard curves were linear over a range of 0.25–50 μg/mL. The precision and accuracy for 0.25 μg/mL (LLQ) in plasma for both compounds were 98.9%, respectively. Interday and intraday precision and accuracy for all QC plasma samples for both compounds were 95.7%, respectively. This methodology details a reproducible assay for both compounds using a single extraction with good accuracy and precision.Keywords
This publication has 4 references indexed in Scilit:
- DEVELOPMENT OF VALIDATED STABILITY INDICATING METHOD BY RP-HPLC FOR SIMULTANEOUS ESTIMATION OF MEROPENEM AND VABORBACTAM IN BULK AND PHARMACEUTICAL FORMULATIONInternational Journal of Pharmacy and Pharmaceutical Sciences, 2019
- A New Simple RP-HPLC Method for Simultaneous Estimation of Meropenem and Vaborbactam in Tablet Dosage FormAsian Journal of Research in Chemistry, 2018
- Pharmacodynamic Assessment of Ertapenem (MK-0826) against Streptococcus pneumoniae in a Murine Neutropenic Thigh Infection ModelAntimicrobial Agents and Chemotherapy, 2002
- High-performance liquid chromatographic assay for meropenem in serumJournal of Chromatography B: Biomedical Sciences and Applications, 1996