Experimental implementation of array‐compressed parallel transmission at 7 tesla
Open Access
- 15 April 2016
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 75 (6) , 2545-2552
Abstract
Purpose: To implement and validate a hardware‐based array‐compressed parallel transmission (acpTx) system.Methods: In array‐compressed parallel transmission, a small number of transmit channels drive a larger number of transmit coils, which are connected via an array compression network that implements optimized coil‐to‐channel combinations. A two channel‐to‐eight coil array compression network was developed using power splitters, attenuators and phase shifters, and a simulation was performed to investigate the effects of coil coupling on power dissipation in a simplified network. An eight coil transmit array was constructed using induced current elimination decoupling, and the coil and network were validated in benchtop measurements, mapping scans, and an accelerated spiral excitation experiment.Results: The developed attenuators came within 0.08 dB of the desired attenuations, and reflection coefficients were −22 dB or better. The simulation demonstrated that up to 3× more power was dissipated in the network when coils were poorly isolated (−9.6 dB), versus well‐isolated (−31 dB). Compared to split circularly‐polarized coil combinations, the additional degrees of freedom provided by the array compression network led to 54% lower squared excitation error in the spiral experiment.Conclusion: Array‐compressed parallel transmission was successfully implemented in a hardware system. Further work is needed to develop remote network tuning and to minimize network power dissipation. Magn Reson Med 75:2545–2552, 2016.Keywords
Funding Information
- National Institutes of Health (R01 EB016695 and R01 DA019912)
This publication has 21 references indexed in Scilit:
- Minimum envelope roughness pulse design for reduced amplifier distortion in parallel excitationMagnetic Resonance in Medicine, 2010
- A 32‐channel lattice transmission line array for parallel transmit and receive MRI at 7 teslaMagnetic Resonance in Medicine, 2010
- Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject‐dependent transmit phase measurementsMagnetic Resonance in Medicine, 2008
- Degenerate mode band‐pass birdcage coil for accelerated parallel excitationMagnetic Resonance in Medicine, 2007
- Spatial domain method for the design of RF pulses in multicoil parallel excitationMagnetic Resonance in Medicine, 2006
- Automatic tuning of flexible interventional RF receiver coilsMagnetic Resonance in Medicine, 2005
- Transmit and receive transmission line arrays for 7 Tesla parallel imagingMagnetic Resonance in Medicine, 2005
- Parallel excitation with an array of transmit coilsMagnetic Resonance in Medicine, 2004
- Transmit SENSEMagnetic Resonance in Medicine, 2002
- Diode Phase Shifters for Array AntennasIEEE Transactions on Microwave Theory and Techniques, 1974