Force-Sensitive Resistors for Detecting Peripheral IV Failure
- 9 September 2025
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Letters
- Vol. 9 (10) , 1-4
Abstract
Peripheral intravenous catheters fail frequently, with some estimates of failure rates as high as 50% globally. In this letter, the potential for using force-sensitive resistors (FSRs) for the detection of swelling caused by infiltration or extravasation associated with peripheral IV failures (PIVIEs) is explored. Silicone models with elastic moduli and shape consistent with that experienced by body tissue during PIVIEs were integrated into a testbed with a layer of artificial skin, transparent film dressing, FSR, and bandage. The response of an FSR manufactured by Tekscan with a standard response range up to 1 lb (4.4 N) was measured for each combination of three swelling heights (8, 12, and 18 mm) and elastic moduli (∼390, ∼490, and ∼540 kPa) and compared to the response of a precise force sensor (SingleTact 1.0 N) to determine the feasibility of using FSRs for detecting PIVIEs. Results show that the FSR resistance consistently decreases with increasing swelling height for a given elastic modulus. The coefficient of variability in the force-sensitive resistor output varies from 0.60% to 31.3% with the lowest variability occurring for the smallest amount of swelling. When filtering is introduced to remove outlying events in the FSR response, however, the range of variability drops to between 0.42% and 10.2%. While these experiments are preliminary, they nevertheless demonstrate the potential for using FSRs in single-use, disposable patches for monitoring IVs and detecting failure due to infiltration and swelling.Keywords
This publication has 22 references indexed in Scilit:
- Piezoresistive relaxation and creep model of porous polymer nanocomposite supported by experimental dataSensors and Actuators A: Physical, 2024
- Multiscale mechanical analysis of the elastic modulus of skinActa Biomaterialia, 2023
- Sensing Technologies for Extravasation Detection: A ReviewACS Sensors, 2023
- Assessment of Skin Properties in Chronic Lymphedema: Measurement of Skin Stiffness, Percentage Water Content, and Transepidermal Water LossLymphatic Research and Biology, 2020
- Ultrasonographic assessment of an induration caused by extravasation of a nonvesicant anticancer drugMedicine, 2019
- The effect of biomechanical variables on force sensitive resistor error: Implications for calibration and improved accuracyJournal of Biomechanics, 2016
- International prevalence of the use of peripheral intravenous cathetersJournal of Hospital Medicine, 2015
- Real Time Monitoring of Arterial Pulse Waveform Parameters Using Low Cost, Non-invasive Force Transducer2015
- Bite Force and Influential Factors on Bite Force Measurements: A Literature ReviewEuropean Journal of Dentistry, 2010
- Time dependence of piezoresistance for the conductor-filled polymer compositesJournal of Polymer Science Part B: Polymer Physics, 2000