Wavefront Automated Refraction Comparison of Three Different IOLs: Aspheric Monofocal and Two Enhanced Monofocal IOLs
Authors:
- Arthur Buffara van Den BergABArthur Buffara van Den Berg
- Roberta Matschinske van Den BergRBRoberta Matschinske van Den Berg
- Bernardo Kaplan MoscoviciBMBernardo Kaplan Moscovici
- Maya DodhiaMDMaya Dodhia
- Larissa GouveaLGLarissa Gouvea
- Wallace ChamonWCWallace Chamon
- Karolinne Maia RochaKRKarolinne Maia Rocha
Open Access
Abstract
The objective of this study was to compare subjective manifest refraction with wavefront-based automated refraction using iTrace (ray tracing) and LadarWave (Hartmann–Shack) in eyes implanted with two enhanced monofocal intraocular lenses (IOLs) and a standard aspheric monofocal IOL, emphasizing agreement and refractive variability across optical designs. This retrospective cohort included 84 eyes from 42 patients implanted with Tecnis Eyhance (DIB00), RayOne EMV (RAO200E), or Tecnis ZCB00 IOLs. Postoperative evaluation (1–3 months) included uncorrected and corrected distance visual acuity and subjective manifest refraction, followed by automated refraction with iTrace and LadarWave. Outcomes were sphere, cylinder, and spherical equivalent (SE). Agreement was assessed using mean signed difference, mean absolute error, root mean square error, Bland–Altman limits of agreement, proportions within clinically relevant thresholds, and vector astigmatism (J0, J45). Linear mixed-effect modeling evaluated SE differences across methods and IOL types while accounting for within-subject correlation. Subjective SE differed among IOLs (p = 0.027), with RAO200E more myopic than ZCB00 (−0.20 ± 0.32 D vs. −0.08 ± 0.44 D, p = 0.035). Automated refraction showed greater variability and poorer agreement in enhanced monofocal IOLs, particularly for cylinder and SE, with wider limits of agreement and fewer eyes within ±0.50 D compared with ZCB00. In mixed-effect contrasts (three-method repeated-measures model), iTrace and LadarWave showed a consistent myopic bias versus manifest refraction in DIB00 and RAO200E, whereas in ZCB00 the iTrace–manifest difference was not significant and LadarWave retained a significant myopic bias. Enhanced monofocal IOLs exhibit reduced agreement between wavefront-based automated and subjective manifest refraction compared with a standard aspheric monofocal IOL. Manifest refraction remains essential for postoperative assessment, and automated measurements should be interpreted as complementary, particularly in IOL designs that modify aberrations.Keywords
This publication has 30 references indexed in Scilit:
- Effect of Monofocal, Wavefront-Shaped, and Diffractive Trifocal Intraocular Lenses on Scanning-Slit Automated RefractionAmerican Journal of Ophthalmology, 2024
- Influence of isofocal intraocular lenses on objective refraction based on autorefraction and aberrometryGraefe's Archive for Clinical and Experimental Ophthalmology, 2023
- Clinical and aberrometric evaluation of a new extended depth-of-focus intraocular lens based on spherical aberrationJournal of Cataract & Refractive Surgery, 2019
- Comparative analysis of visual outcomes with 4 intraocular lenses: Monofocal, multifocal, and extended range of visionJournal of Cataract & Refractive Surgery, 2018
- Evaluation of visual outcomes and patient satisfaction after implantation of a diffractive trifocal intraocular lensJournal of Cataract & Refractive Surgery, 2016
- Ocular wavefront analysis of aspheric compared with spherical monofocal intraocular lenses in cataract surgery: Systematic review with metaanalysisJournal of Cataract & Refractive Surgery, 2015
- In vitro antimicrobial analysis of aqueous humor after topical application of moxifloxacin hydrochloride 0.5%Journal of Cataract & Refractive Surgery, 2015
- Multifocal Intraocular Lens Explantation: A Case Series of 50 EyesAmerican Journal of Ophthalmology, 2014
- Autorefraction after implantation of diffractive multifocal intraocular lensesJournal of Cataract & Refractive Surgery, 2010
- Expanding depth of focus by modifying higher-order aberrations induced by an adaptive optics visual simulatorJournal of Cataract & Refractive Surgery, 2009