Reliability of Evoked Endocardial T‐wave Sensing in 1,500 Pacemaker Patients
Authors:
- 1 November 1986
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 9 (6) , 948-953
Abstract
Intracardiac measurements of the evoked endocardial T-wave were performed on 1,500 pacemaker patients. The sensed evoked T-wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R-wave and that of the evoked T-wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2, porous surface structure (preferably carbon), and atraumatic fixation.Keywords