AC/DC coupling effects on CW and pulse transmission modes in Doppler radar physiological monitoring system
Authors:
- 1 January 2012
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 25-28
Abstract
Direct-conversion CW microwave Doppler radar can be used to wirelessly detect cardiopulmonary activity. One of the limitations of homodyne CW Doppler radar systems for physiological monitoring is large DC offset in baseband outputs. The common method to avoid the DC offset is AC coupling. While AC coupling removes the DC offset efficiently, it introduces large settling time and signal distortion in baseband. In this paper we explore the use of direction conversion pulsed Doppler radar to overcome this issue. Performance of CWand pulse radar is compared using mechanical target movement which simulates respiratory motion. The results demonstrate while AC coupling distorts CW radar output, it has a negligible effect on pulse radar output.Keywords
This publication has 4 references indexed in Scilit:
- System-on-Chip based Doppler radar occupancy sensor2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2011
- DC information preservation for cardiopulmonary monitor utilizing CW doppler radar2008
- Arctangent Demodulation With DC Offset Compensation in Quadrature Doppler Radar Receiver SystemsIEEE Transactions on Microwave Theory and Techniques, 2007
- Range Correlation and$ I/ Q$Performance Benefits in Single-Chip Silicon Doppler Radars for Noncontact Cardiopulmonary MonitoringIEEE Transactions on Microwave Theory and Techniques, 2004