A study on the theoretical and practical accuracy of conoscopic holography‐based surface measurements: toward image registration in minimally invasive surgery

Authors:

Abstract
Background: Registered medical images can assist with surgical navigation and enable image‐guided therapy delivery. In soft tissues, surface‐based registration is often used and can be facilitated by laser surface scanning. Tracked conoscopic holography (which provides distance measurements) has been recently proposed as a minimally invasive way to obtain surface scans. Moving this technique from concept to clinical use requires a rigorous accuracy evaluation, which is the purpose of our paper.Methods: We adapt recent non‐homogeneous and anisotropic point‐based registration results to provide a theoretical framework for predicting the accuracy of tracked distance measurement systems. Experiments are conducted a complex objects of defined geometry, an anthropomorphic kidney phantom and a human cadaver kidney.Results: Experiments agree with model predictions, producing point RMS errors consistently < 1 mm, surface‐based registration with mean closest point error < 1 mm in the phantom and a RMS target registration error of 0.8 mm in the human cadaver kidney.Conclusions: Tracked conoscopic holography is clinically viable; it enables minimally invasive surface scan accuracy comparable to current clinical methods that require open surgery. Copyright © 2012 John Wiley & Sons, Ltd.