Multipoint temperature control during hyperthermia treatments: theory and simulation
Authors:
- 1 January 1995
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 42 (8) , 818-827
Abstract
A real-time multipoint feedback temperature control system has been designed and implemented with an ultrasound phased-array applicator for hyperthermia. The control parameters are the total power available from the supply and the dwell times at a sequence of preselected heating patterns. Thermocouple measurements are assumed for temperature feedback. The spatial operator linking available heating patterns to temperature measurements is measured at the outset of the treatment and can be remeasured on line an adaptive implementation. A significant advantage of this approach is that the controller does not require a priori knowledge of either the placement of the thermocouples or the power distribution of the ultrasound heating patterns. Furthermore, the control loop uses a proportional integral (PI) gain in conjunction with a singular value decomposition (SVD) of the spatial transfer operator. This approach is advantageous for robust implementation and is shown to properly balance the power applied to the individual patterns. The controller also deals with saturation in the inputs without integrator windup and, therefore, without temperature overshoot. In this paper, we present the theoretical formulation and representative simulation results of the proposed controller. The control algorithm has been verified experimentally, both in vitro and in vivo. A subsequent paper describing these results and the practical implementation of the controller will follow.Keywords
This publication has 22 references indexed in Scilit:
- Multipoint temperature control during hyperthermia treatments: theory and simulationIEEE Transactions on Biomedical Engineering, 1995
- A new algorithm for dynamic focusing of phased-array hyperthermia applicators through tissue inhomogeneities1993
- Performance of an adaptive MIMO controller for a multiple-element ultrasound hyperthermia systemInternational Journal of Hyperthermia, 1993
- Theoretical and experimental evaluation of a temperature controller for scanned focused ultrasound hyperthermiaMedical Physics, 1990
- Multi-point feedback control system for scanned, focused ultrasound hyperthermiaPhysics in Medicine & Biology, 1990
- Temperature artifacts produced by thermocouples used in conjunction with 1 and 3 MHz ultrasoundInternational Journal of Hyperthermia, 1990
- Catheter induced temperature artifacts in ultrasound hyperthermiaInternational Journal of Hyperthermia, 1990
- Invited review: Noninvasive microwave phased arrays for local hyperthermia: A reviewInternational Journal of Hyperthermia, 1989
- Simulation of automatic temperature control in tissue hyperthermia calculationsMedical Physics, 1985
- On the input-output stability of time-varying nonlinear feedback systems Part one: Conditions derived using concepts of loop gain, conicity, and positivityIEEE Transactions on Automatic Control, 1966