Real-Time Control of Focused Ultrasound Heating Based on Rapid MR Thermometry

Authors:
  • JZJacco A. DE Zwart
    Jacco A. DE Zwart
  • JPJean Palussiére
    Jean Palussiére
  • RFRabia Fawaz
    Rabia Fawaz
  • CDChristophe Delalande
    Christophe Delalande
  • PCPaul Canioni
    Paul Canioni
  • NGNicolas Grenier
    Nicolas Grenier
  • CMChrit T. W. Moonen
    Chrit T. W. Moonen

Abstract
Real-time control of the heating procedure is essential for hyperthermia applications of focused ultrasound (FUS). The objective of this study is to demonstrate the feasibility of MRI-controlled FUS. An automatic control system was developed using a dedicated interface between the MR system control computer and the FUS wave generator. Two algorithms were used to regulate FUS power to maintain the focal point temperature at a desired level. Automatic control of FUS power level was demonstrated ex vivo at three target temperature levels (increase of 5° C, 10° C, and 30° C above room temperature) during 30-minute hyperthermic periods. Preliminary in vivo results on rat leg muscle confirm that necrosis estimate, calculated on-line during FUS sonication, allows prediction of tissue damage. The feasibility of fully automatic FUS control based on MRI thermometry has been demonstrated.