For many years the free-air ionization chamber has been the standard instrument for the measurement in roentgens of x-radiation doses below 200 kv. Agreement between the standard chambers of several national laboratories within about ±1 per cent was established in 1931 by Taylor (1, 2), and the design of the standard chamber (3, 4) at the National Bureau of Standards has remained unchanged from that time until recently. In 1952, Kemp and Hall (5) carried out a precise study of the lateral dimension requirements for complete utilization of electron ionization in a parallel-plate, variable-pressure air chamber. Their results indicated that the standard chambers at both the National Bureau of Standards (NBS) and the National Physical Laboratory (NPL) in England were inadequate in this respect, and that electric field distortion was responsible for the consistent under-estimation of plate separation requirements by other workers. Preliminary measurements carried out in this laboratory early in 1953, with a la...