Genetic algorithm search for the worst‐case MRI RF exposure for a multiconfiguration implantable fixation system modeled using artificial neural networks
- 27 May 2020
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 84 (5) , 2754-2764
Abstract
Purpose: This paper presents a method to search for the worst‐case configuration leading to the highest RF exposure for a multiconfiguration implantable fixation system under MRI.Methods: A two‐step method combining an artificial neural network and a genetic algorithm is developed to achieve this purpose. In the first step, the level of RF exposure in terms of peak 1‐g and/or 10‐g averaged specific absorption rate (SAR1g/10g), related to the multiconfiguration system, is predicted using an artificial neural network. A genetic algorithm is then used to search for the worst‐case configuration of this multidimensional nonlinear problem within both the enumerated discrete sample space and generalized continuous sample space. As an example, a generic plate system with a total of 576 configurations is used for both 1.5T and 3T MRI systems.Results: The presented method can effectively identify the worst‐case configuration and accurately predict the SAR1g/10g with no more than 20% of the samples in the studied discrete sample space, and can even predict the worst case in the generalized continuous sample space. The worst‐case prediction error in the generalized continuous sample space is less than 1.6% for SAR1g and less than 1.3% for SAR10g compared with the simulation results.Conclusion: The combination of an artificial neural network with genetic algorithm is a robust technique to determine the worst‐case RF exposure level for a multiconfiguration system, and only needs a small amount of training data from the entire system.Keywords
Funding Information
- National Science Foundation (1440107)
This publication has 14 references indexed in Scilit:
- Prediction of MRI RF Exposure for Implantable Plate Devices Using Artificial Neural NetworkIEEE Transactions on Electromagnetic Compatibility, 2019
- Retrospective analysis of RF heating measurements of passive medical implantsMagnetic Resonance in Medicine, 2018
- Impacts of RF shimming on local SAR caused by MRI 3T birdcage coil near femoral plate implants2017
- Parallel radiofrequency transmission at 3 tesla to improve safety in bilateral implanted wires in a heterogeneous modelMagnetic Resonance in Medicine, 2017
- Numerical study of SAR for multi-component orthopaedic hip replacement system during MRI2016
- Comparison of simulated parallel transmit body arrays at 3 T using excitation uniformity, global SAR, local SAR, and power efficiency metricsMagnetic Resonance in Medicine, 2014
- Impact of capped and uncapped abandoned leads on the heating of an MR‐conditional pacemaker implantMagnetic Resonance in Medicine, 2014
- Thermal Tissue Damage Model Analyzed for Different Whole‐Body SAR and Scan Durations for Standard MR Body CoilsMagnetic Resonance in Medicine, 2013
- MRI estimation of contrast agent concentration in tissue using a neural network approachMagnetic Resonance in Medicine, 2007
- Prepolarized magnetic resonance imaging around metal orthopedic implantsMagnetic Resonance in Medicine, 2006