We describe here the increased yield of dielectrophoretic capture of nucleic acids as a function of modifications to the sample. To this end, we investigated the effect of different buffer formulations in the capture of λ-DNA in interdigitated electrodes. We report an initial assessment of the effect of changes in conductivity and viscosity, as well as the specific role of different organic solvents and Mg2+. Our experiments demonstrated that the controlled addition of acetonitrile can significantly improve the capture of λ-DNA. This might open the path to capture fragmented nucleic acids, such as cell-free DNA, enabling improved sensing and quantification, and enabling applications such as liquid biopsy.