High-Field-Strength MR Imaging of the Liver at 3.0 T: Intraindividual Comparative Study with MR Imaging at 1.5 T
Authors:
- Marcus M. von FalkenhausenMFMarcus M. von Falkenhausen
- Götz LutterbeyGLGötz Lutterbey
- Nuschin Morakkabati-SpitzNMNuschin Morakkabati-Spitz
- Oliver WalterOWOliver Walter
- Jürgen GiesekeJGJürgen Gieseke
- Renate BlömerRBRenate Blömer
- Winfried A. WillinekWWWinfried A. Willinek
- Hans H. SchildHSHans H. Schild
- Christiane K. KuhlCKChristiane K. Kuhl
- 1 October 2006
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 241 (1) , 156-166
Abstract
To prospectively evaluate whether magnetic resonance (MR) imaging of the liver at 3.0 T is comparable to that at 1.5 T with respect to image artifacts, image quality, and diagnostic utility in terms of detection and characterization of focal liver lesions in patients with these lesions. Patients provided informed consent after the study had been explained, and the institutional review board approved the study protocol. An intraindividual comparative study was performed in 21 patients (12 men and nine women; mean age, 58.7 years; range, 36-76 years) with a total of 79 focal liver lesions (benign and malignant) who were examined at 1.5- and 3.0-T MR imaging within 1 week. The imaging protocol consisted of T2-weighted turbo spin-echo (SE) sequences with or without fat suppression, as well as T1-weighted gradient-echo (GRE) sequences with or without gadolinium-based contrast agent. All images were rated independently by two radiologists with respect to types of artifacts (susceptibility, motion, pulsation, image homogeneity, and electrodynamic effects) and in regard to detectability and characterization of focal liver lesions. A modified sign test was used for statistical analysis (alpha < .2). Motion artifacts were significantly more pronounced in non-fat-suppressed T2-weighted turbo SE images at 3.0 T (P = .03), whereas pulsation artifacts were more pronounced (P = .19) in precontrast T1-weighted GRE 1.5-T images. No statistically significant differences (P < .2) were observed for the remaining artifacts and sequences. Of the 79 index lesions, a total of 76 were prospectively identified at 1.5-T imaging and a total of 77 were identified at 3.0-T imaging. MR imaging of the liver at 3.0 T, compared with that at 1.5 T, is feasible with equivalent image quality and diagnostic utility in terms of detection and characterization of focal liver lesions.Keywords
This publication has 17 references indexed in Scilit:
- Functional 3.0-T MR Assessment of Higher Cognitive Function: Are There Advantages over 1.5-T Imaging?Radiology, 2005
- Acute and Subacute Ischemic Stroke at High-Field-Strength (3.0-T) Diffusion-weighted MR Imaging: Intraindividual Comparative StudyRadiology, 2005
- Whole-body MRI at high field: technical limits and clinical potentialEuropean Radiology, 2005
- Klinische MRT bei 3 Tesla: Aktueller StandDer Radiologe, 2004
- Time-of-Flight MR Angiography: Comparison of 3.0-T Imaging and 1.5-T Imaging—Initial ExperienceRadiology, 2003
- Clinical Experience on 3.0 T Systems in Niigata, 1996 to 2002Investigative Radiology, 2003
- Magnetic Resonance Imaging at 3.0 Tesla: Challenges and Advantages in Clinical Neurological ImagingInvestigative Radiology, 2003
- Criteria for Normalcy of Cavities Observed Within the Adult Hippocampus: High-Resolution Magnetic Resonance Imaging Study on a 3.0-T SystemJournal of Neuroimaging, 2002
- Neuroimaging at 1.5 T and 3.0 T: Comparison of oxygenation‐sensitive magnetic resonance imagingMagnetic Resonance in Medicine, 2001
- Standing-wave and RF penetration artifacts caused by elliptic geometry: an electrodynamic analysis of MRIIEEE Transactions on Medical Imaging, 1998