Impact of arterial location, pressure wave indicators, and measurement devices on arterial form factor and mean and central arterial pressure
Authors:
- 9 December 2022
- journal article
- research article
- Published by Springer Nature in Journal of Human Hypertension
- Vol. 37 (10) , 891-897
Abstract
Mean arterial pressure (MAP) is often estimated from cuff systolic (S) and diastolic (D) blood pressure (BP) using a fixed arterial form factor (FF, usually 0.33). If MAP is measured directly, a true FF can be calculated: FF = [MAP–DBP]/[SBP–DBP]. Because waveform shapes vary, true FF should also vary and MAP accuracy will be affected. We studied factors affecting FF using radial tonography (SphygmoCor, n = 376) or brachial oscillometry (Mobil-O-Graph, n = 157) and to compare devices, 101 pairs were matched precisely for SBP and DBP. SphygmoCor brachioradial FF correlated strongly with central FF (r2 = 0.75), central augmentation index (cAI, r2 = 0.39), and inversely with pulse pressure amplification (PPA) ratio (r2 = 0.44) [all p < 0.000]; brachioradial FF was lower than central (c) FF (0.34 vs. 0.44, 95% CI’s [0.23,0.46] and [0.34,0.54], p < 0.000). On forward stepwise regression, brachioradial FF correlated with PPA ratio, age, heart rate, and cAI (multiple-r2 0.63, p < 0.000). With Mobil-O-Graph: brachial FF was fixed, lower than the corresponding cFF [mean(SD)] 0.46(0.00098) vs. 0.57(0.048), p < 0.000], and uncorrelated with clinical characteristics; MAP and cSBP were higher than SphygmoCor by 6.3 and 2.2 mmHg (p < 0.005) at the midpoint with systematic negative biases. We conclude that FF derived from radial tonometry (SphygmoCor) varies with pulse wave morphology within and between individuals and by measurement site, age, and heart rate. With oscillometry (Mobil-O-Graph), brachial FF was fixed and high and unrelated to other clinical variables; MAP and cSBP were higher than tonometry, with systematic negative biases.Keywords
All Related Versions
This publication has 28 references indexed in Scilit:
- Hemodynamic Correlates of Blood Pressure Across the Adult Age SpectrumCirculation, 2010
- A new oscillometric method for pulse wave analysis: comparison with a common tonometric methodJournal of Human Hypertension, 2010
- Augmentation Index, Left Ventricular Contractility, and Wave ReflectionHypertension, 2009
- Amplification of the Pressure Pulse in the Upper Limb in Healthy, Middle-Aged Men and WomenHypertension, 2009
- Noninvasive Assessment of Local Pulse PressureHypertension, 2005
- Changes in the derived central pressure waveform and pulse pressure in response to angiotensin II and noradrenaline in manThe Journal of Physiology, 2001
- A Comparison of the Indirect Estimate of Mean Arterial Pressure Calculated by the Conventional Equation and Calculated to Compensate for a Change in Heart RateInternational Journal of Sports Medicine, 2000
- Short- and Long-term Effects of Antihypertensive Drugs on Arterial Reflections, Compliance, and ImpedanceHypertension, 1995
- Mechanics of the occlusive arm cuff and its application as a volume sensorIEEE Transactions on Biomedical Engineering, 1993
- An analysis of the relationship between central aortic and peripheral upper limb pressure waves in manEuropean Heart Journal, 1993