New optically stimulated luminescence dosimetry film optimized for energy dependence guided by Monte Carlo simulations
- 22 March 2024
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 69 (7) , 075005
Abstract
Optically stimulated luminescence (OSL) film dosimeters, based on BaFBr:Eu 2+ phosphor material, have major dosimetric advantages such as dose linearity, high spatial resolution, film re-usability, and immediate film readout. However, they exhibit an energy-dependent over-response at low photon energies because they are not made of tissue-equivalent materials. In this work, the OSL energy-dependent response was optimized by lowering the phosphor grain size and seeking an optimal choice of phosphor concentration and film thickness to achieve sufficient signal sensitivity. This optimization process combines measurement-based assessments of energy response in narrow x-ray beams with various energy response calculation methods applied to different film metrics. Theoretical approaches and MC dose simulations were used for homogeneous phosphor distributions and for isolated phosphor grains of different dimensions, where the dose in the phosphor grain was calculated. In total 8 OSL films were manufactured with different BaFBr:Eu 2+ median particle diameters (D50): 3.2 μ m, 1.5 μ m and 230 nm and different phosphor concentrations (1.6%, 5.3% and 21.3 %) and thicknesses (from 5.2 to 49 μ m). Films were irradiated in narrow x-ray spectra (N60, N80, N-150 and N-300) and the signal intensity relative to the nominal dose-to-water value was normalized to Co-60. Finally, we experimentally tested the response of several films in Varian 6MV TrueBeam STx linear accelerator using the following settings: 10 × 10 cm 2 field, 0deggantry angle, 90 cm SSD, 10 cm depth. The x-ray irradiation experiment reported a reduced energy response for the smallest grain size with an inverse correlation between response and grain size. The N-60 irradiation showed a 43% reduction in the energy over-response when going from 3 μ m to 230 nm grain size for the 5% phosphor concentration. Energy response calculation using a homogeneous dispersion of the phosphor underestimated the experimental response and was not able to obtain the experimental correlation between grain size and energy response. Isolated grain size modeling combined with MC dose simulations allowed to establish a good agreement with experimental data, and enabled steering the production of optimized OSL-films. The clinical 6 MV beam test confirmed a reduction in energy dependence, which is visible in small-grain films where a decrease in out-of-field over-response was observed.Keywords
Funding Information
- Agentschap Innoveren & Ondernemen (HBC.2020.3003)
This publication has 45 references indexed in Scilit:
- Optically stimulated luminescence system as an alternative for radiochromic film for 2D reference dosimetry in UHDR electron beamsPhysica Medica, 2023
- Calibration and time fading characterization of a new optically stimulated luminescence film dosimeterMedical Physics, 2022
- Investigation of 2D radiophotoluminescence films in radiotherapy: Multileaf collimator Quality Assurance and small field dosimetryNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2022
- Basics of Radiation Therapy2020
- Practical application of Octavius®‐4D: Characteristics and criticalities for IMRT and VMAT verificationJournal of Applied Clinical Medical Physics, 2018
- Tolerance limits and methodologies forIMRTmeasurement‐based verificationQA:Recommendations ofAAPMTask Group No. 218Medical Physics, 2018
- Development of a 2D dosimetry system based on the optically stimulated luminescence of Al2O3Radiation Measurements, 2014
- Clinical tests of large area thermoluminescent detectors under radiotherapy beamsRadiation Measurements, 2013
- 2D-OSL-dosimetry with beryllium oxideRadiation Measurements, 2010
- Scanning orientation effects on Gafchromic EBT film dosimetryAustralasian Physics & Engineering Sciences in Medicine, 2006