Physical characterization of a scanning photon counting digital mammography system based on Si‐strip detectors
Authors:
- 8 May 2007
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (6Part1) , 1918-1925
Abstract
The physical performance of a scanning multislit full field digital mammography system was determined using basic image quality parameters. The system employs a direct detection detector comprised of linear silicon strip sensors in an edge‐on geometry connected to photon counting electronics. The pixel size is and the field of view . The performance was quantified using the presampled modulation transfer function, the normalized noise power spectrum and the detective quantum efficiency (DQE). Compared to conventional DQE methods, the scanning geometry with its intrinsic scatter rejection poses additional requirements on the measurement setup, which are investigated in this work. The DQE of the photon counting system was found to be independent of the dose level to the detector in the range. The peak DQE was 72% and 73% in the scan and slit direction, respectively, measured with a anode‐filter combination with an added filtration.Keywords
This publication has 22 references indexed in Scilit:
- Scatter rejection in multislit digital mammographyMedical Physics, 2006
- Physical characterization of a prototype selenium-based full field digital mammography detectorMedical Physics, 2005
- System performance of a prototype flat‐panel imager operated under mammographic conditionsMedical Physics, 2003
- Performance evaluation of a “dual‐side read” dedicated mammography computed radiography systemMedical Physics, 2003
- Imaging performance of amorphous selenium based flat-panel detectors for digital mammography: Characterization of a small area prototype detectorMedical Physics, 2003
- ATLAS silicon microstrip Semiconductor Tracker (SCT)Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2000
- Analysis of the detective quantum efficiency of a developmental detector for digital mammographyMedical Physics, 1999
- Signal-to-noise optimization of medical imaging systemsJournal of the Optical Society of America A, 1999
- Scanned‐projection digital mammographyMedical Physics, 1987
- The Equivalent Quantum Efficiency of the Photographic ProcessThe Journal of Photographic Science, 1963