Abstract
An automatic fluorimetric method for the determination of boron in water samples and soil extracts is described. Automatic wet digestion of the samples by UV irradiation, eliminates interference by organic matter. An ion exchange resin incorporated in the system removed interference caused by metal cations. The analysis is performed at pH 7. The linear working range is 0 ‐ 3 mg 1−1 of boron with a detection limit of 0.01 mg 1−1 of boron. The capacity of the method is a minimum of 20 samples per hour and the precision is s = 0.004 mg 1−1 of boron in the range 0 ‐ 3 mg 1−1 of boron.

This publication has 7 references indexed in Scilit: