Convex optimization of MRI exposure for mitigation of RF-heating from active medical implants
- 9 September 2015
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 60 (18) , 7293-7308
Abstract
Local RF-heating of elongated medical implants during magnetic resonance imaging (MRI) may pose a significant health risk to patients. The actual patient risk depends on various parameters including RF magnetic field strength and frequency, MR coil design, patient’s anatomy, posture, and imaging position, implant location, RF coupling efficiency of the implant, and the bio-physiological responses associated with the induced local heating. We present three constrained convex optimization strategies that incorporate the implant’s RF-heating characteristics, for the reduction of local heating of medical implants during MRI. The study emphasizes the complementary performances of the different formulations. The analysis demonstrates that RF-induced heating of elongated metallic medical implants can be carefully controlled and balanced against MRI quality. A reduction of heating of up to 25 dB can be achieved at the cost of reduced uniformity in the magnitude of the field of less than 5%. The current formulations incorporate a priori knowledge of clinically-specific parameters, which is assumed to be available. Before these techniques can be applied practically in the broader clinical context, further investigations are needed to determine whether reduced access to a priori knowledge regarding, e.g. the patient’s anatomy, implant routing, RF-transmitter, and RF-implant coupling, can be accepted within reasonable levels of uncertainty.Keywords
Funding Information
- Ministry of Education, Spain (Jose Castillejo Research Mobility Programme) (CAS12/00216)
This publication has 27 references indexed in Scilit:
- Studies of RF shimming techniques with minimization of RF power deposition and their associated temperature changesConcepts in Magnetic Resonance Part B, 2011
- FIELD SYNTHESIS IN INHOMOGENEOUS MEDIA: JOINT CONTROL OF POLARIZATION, UNIFORMITY AND SAR IN MRI B1-FIELDProgress In Electromagnetics Research, 2011
- The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulationsPhysics in Medicine & Biology, 2009
- $B_1$ Homogenization in MRI by Multilayer Coupled CoilsIEEE Transactions on Medical Imaging, 2008
- Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 Tesla with eight channelsMagnetic Resonance in Medicine, 2008
- Calculation of MRI‐induced heating of an implanted medical lead wire with an electric field transfer functionJournal of Magnetic Resonance Imaging, 2007
- 7 T body MRI:B1shimming with simultaneous SAR reductionPhysics in Medicine & Biology, 2007
- Transmission line for improved RF safety of interventional devicesMagnetic Resonance in Medicine, 2005
- In vivo heating of pacemaker leads during magnetic resonance imagingEuropean Heart Journal, 2004
- Transient dystonia following magnetic resonance imaging in a patient with deep brain stimulation electrodes for the treatment of Parkinson diseaseJournal of Neurosurgery, 2003