Resistance valves in circulatory loops have a significant impact on in vitro evaluation of blood damage caused by blood pumps: a computational study
Open Access
- 30 November 2023
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Physiology
- Vol. 14, 1287207
Abstract
Background: Hemolysis and its complications are major concerns during the clinical application of blood pumps. In-vitro circulatory testing loops have been employed as the key procedure to evaluate the hemolytic and thrombogenic performance of blood pumps during the development phase and before preclinical in-vivo animal studies. Except for the blood damage induced by the pump under test, blood damage induced by loop components such as the resistance valve may affect the accuracy, reproducibility, and intercomparability of test results.Methods: This study quantitatively investigated the impact of the resistance valve on in vitro evaluation of blood damage caused by blood pumps under different operating points. A series of idealized tubing models under the resistance valve with different openings were created. Three pumps - the FDA benchmark pump, the HeartMate 3 LVAD, and the CH-VAD - were involved in hypothetical tests. Eight operating points were chosen to cover a relatively wide spectrum of testing scenarios. Computational fluid dynamics (CFD) simulations of the tubing and pump models were conducted at the same operating points.Results and Conclusion: Overall, hemolysis and platelet activation induced by a typical resistance valve are equivalent to 17%–45% and 14%–60%, respectively, of those induced by the pump itself. Both ratios varied greatly with flow rate, valve opening and pump models. Differences in blood damage levels between different blood pumps or working conditions can be attenuated by up to 45%. Thus, hemolysis and platelet activation induced by the resistance valve significantly affect the accuracy of in-vitro hemocompatibility evaluations of blood pumps. A more accurate and credible method for hemocompatibility evaluations of blood pumps will benefit from these findings.Keywords
This publication has 37 references indexed in Scilit:
- Quantification of Shear‐Induced Platelet Activation: High Shear Stresses for Short Exposure TimeArtificial Organs, 2015
- Multilaboratory Study of Flow-Induced Hemolysis Using the FDA Benchmark Nozzle ModelArtificial Organs, 2014
- Ambient hemolysis and activation of coagulation is different between HeartMate II and HeartWare left ventricular assist devicesThe Journal of Heart and Lung Transplantation, 2014
- Study of Flow-Induced Hemolysis Using Novel Couette-Type Blood-Shearing DevicesArtificial Organs, 2011
- Significance of Extensional Stresses to Red Blood Cell Lysis in a Shearing FlowAnnals of Biomedical Engineering, 2011
- The use of computational fluid dynamics in the development of ventricular assist devicesMedical Engineering & Physics, 2010
- On the representation of turbulent stresses for computing blood damageInternational Journal of Engineering Science, 2010
- Fast Three‐dimensional Numerical Hemolysis ApproximationArtificial Organs, 2004
- Mechanical Blood Traumatization by Tubing and Throttles in In Vitro Pump Tests: Experimental Results and Implications for Hemolysis TheoryArtificial Organs, 1993
- Hemolytic effects of energy dissipation in flowing bloodMedical & Biological Engineering & Computing, 1969