Nitrogen ion implantation on a Ti6Al4V alloy with 70KeV energy was carried out at different doses ranging from 5 × 1015to 2.5 × 1017ions/cm2. The implanted samples were subjected to open circuit potential/time measurement and cyclic polarization studies to evolve the optimum dose which can give good corrosion resistance in a simulated body fluid condition. The results show that there was an increase in corrosion resistance with increased doses up to 7 × 1016ions/cm2, beyond which it started to decrease. The stability of the passive film at a higher potential was assessed by potentiotransient techniques after impressing a constant potential of 1.5V for three hours. The results of the investigation indicate that nitrogen ion implantation can be used as a viable method to improve the corrosion resistance of orthopaedic implant devices made of Ti6Al4V alloy. The nature of the stable passive film and its influence on corrosion resistance are discussed in this paper.Keywords
This publication has 18 references indexed in Scilit:
- Dissolution, Passivation and Composition of Passive Films of Binary Iron-Nitrogen AlloyMaterials Science Forum, 1995
- Corrosion fatigue resistances of surgical implant stainless steels and titanium alloyCorrosion Science, 1993
- Tissue-biomaterial interactionsJournal of Materials Science, 1987
- In vivo metal‐ion release from porous titanium‐fiber materialJournal of Biomedical Materials Research, 1984
- Surface Enrichment of Nitrogen during Passivation of a Highly Resistant Stainless SteelJournal of the Electrochemical Society, 1983
- In vivo and in vitro studies of the stress‐corrosion cracking behavior of surgical implant alloysJournal of Biomedical Materials Research, 1983
- Instability of Anodic Oxide Films on TitaniumJournal of the Electrochemical Society, 1982
- Some new studies of the wear behavior of ultrahigh molecular weight polyethyleneJournal of Biomedical Materials Research, 1976
- The application of ion beams to corrosion science☆Corrosion Science, 1976
- Changes in nonosseous tissue adjacent to titanium implantsJournal of Biomedical Materials Research, 1973