Multicenter Study of the Electrical Characteristics and Short‐Term Outcomes of the Aveir VR Leadless Pacemaker
- 6 October 2026
- journal article
- Published by Wolters Kluwer Health in Journal of the American Heart Association
- Vol. 15 (19) , e050579
Abstract
Background: The Aveir leadless pacemaker uses an active fixation method, enabling real‐time monitoring of electrical parameters during implantation. However, comprehensive studies on the electrical parameters during this procedure are rare. This study aims to analyze the electrical characteristics to further guide the implantation strategy and improve device stability and safety. Methods: This multicenter retrospective study enrolled 119 patients (mean age, 70.18 years; 59.58% women) who received the Aveir VR leadless pacemaker from November 2024 to May 2025 at 10 centers in China. Intraprocedural variations in commanded electrogram, current of injury (COI), impedance, pacing threshold, and sensing parameters were meticulously documented. Results: Commanded electrogram mapping demonstrated various morphologies (R, RS, QR, QRS, and QS), aiding localization. During fixation, 58.82% of patients exhibited an increased COI from mapping to 0.5 turns, which was associated with reduced short‐term pacing thresholds. From 0.5 to 1 turn, 52.94% showed a further increase in COI. Receiver operating characteristic analysis revealed that an impedance increase has predictive value for short‐term pacing thresholds, with an area under the receiver operating characteristic curve of 0.634 and a cutoff value of 230 Ω (sensitivity, 0.622; specificity, 0.41). Lead stability showed a moderate correlation with impedance increase ( R =0.44, P <0.001), while the correlation with COI was weak. Conclusions: During Aveir implantation, commanded electrogram variations guide site localization. Initial COI increases (0–0.5 turns) are linked to optimal short‐term thresholds. Monitoring impedance increase is vital, as a threshold of 230 Ω serves as a key indicator of device stability and fixation quality.Keywords
This publication has 10 references indexed in Scilit:
- Aveir VR real-world performance and chronic pacing threshold prediction using mapping and fixation electrical dataEP Europace, 2024
- A Dual-Chamber Leadless PacemakerNew England Journal of Medicine, 2023
- Linee guida ESC 2021 su stimolazione cardiaca e terapia di resincronizzazione cardiaca elaborate dalla Task Force su stimolazione cardiaca e terapia di resincronizzazione cardiaca della Società Europea di Cardiologia (ESC) con il contributo straordinario della European Heart Rhythm Association (EHRA)Giornale Italiano di Cardiologia, 2022
- Current of injury is an indicator of lead depth and performance during left bundle branch pacing lead implantationHeart Rhythm, 2022
- Primary Results on Safety and Efficacy From the LEADLESS II–Phase 2 Worldwide Clinical TrialJACC: Clinical Electrophysiology, 2022
- Leadless pacemaker versus transvenous single-chamber pacemaker therapy: A propensity score-matched analysisHeart Rhythm, 2018
- Human ventricular activation sequence and the simulation of the electrocardiographic QRS complex and its variability in healthy and intraventricular block conditionsEP Europace, 2016
- A Leadless Intracardiac Transcatheter Pacing SystemNew England Journal of Medicine, 2016
- 50th Anniversary of the First Successful Permanent Pacemaker Implantation in the United States: Historical Review and Future DirectionsThe American Journal of Cardiology, 2010
- Current of injury predicts adequate active lead fixation in permanent pacemaker/defibrillation leadsJACC, 2005