Operation of a Miniature Redox Hydrogel-Based Pyruvate Sensor in Undiluted Deoxygenated Calf Serum
Authors:
- 23 May 2000
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 72 (13) , 2963-2968
Abstract
An amperometric sensor for the detection of pyruvate in biological fluids was formed by modifying the tip of a 0.25-mm gold wire with a layer of electrically “wired” recombinant pyruvate oxidase (POP). The sensor did not require O2 for its operation. The electroactive area of the tip of the microwire was increased by electrodeposition of platinum black. The POP was adsorbed on the platinum black and then “wired” with the cross-linked, subsequently deposited poly(4-vinylpyridine), part of the pyridine functions of which were complexed with [Os(bpy)2Cl]+/2+ and part quaternized with 2-bromoethylamine. In the resulting thin layer the POP was well “wired”. When the electrode was poised at 0.4 V vs Ag/AgCl, the sensitivity at pH 6 was 0.26 A cm-2 M-1 and the current increased linearly with the pyruvate concentration through the 2 × 10-6−6 × 10-4 M range. Thiamine diphosphate, flavin adenine dinucleotide, and MgCl2 were not required for the assay, but stabilized the stored enzyme electrode. Placement of a dialysis membrane (MWCO 3500) on the electrode alleviated the severe interference of ascorbate. In calf serum, the detection limit was 30 μM, suggesting that the electrode might be used in the continuous monitoring of pyruvate in hypoxic organs.Keywords
This publication has 12 references indexed in Scilit:
- Pyruvate sensor based on pyruvate oxidase immobilized in a poly(mercapto-p-benzoquinone) filmJournal of Electroanalytical Chemistry, 1999
- Microbial cyanide sensor for monitoring river waterJournal of Biotechnology, 1996
- Effect of Quaternization of the Glucose Oxidase “Wiring” Redox Polymer on the Maximum Current Densities of Glucose ElectrodesThe Journal of Physical Chemistry, 1996
- Carbon paste electrodes modified with enzymes, tissues, and cellsElectroanalysis, 1995
- Electrostatic Control of the Electron-Transfer Enabling Binding of Recombinant Glucose Oxidase and Redox PolyelectrolytesJournal of the American Chemical Society, 1994
- Amperometric methylene green-mediated pyruvate electrode based on pyruvate oxidase entrapped in carbon pasteAnalytica Chimica Acta, 1992
- Phosphate Sensor Composed from Immobilized Pyruvate Oxidase and an Oxygen ElectrodeAnalytical Letters, 1991
- Enzyme electrode composed of the pyruvate oxidase from pediococcus species coupled to an oxygen electrode for measurements of pyruvate in biological mediaBiosensors, 1987
- Sequential determination of glutamate-oxalacetate transaminase and glutamate-pyruvate transaminase activities in serum using an immobilized bienzyme-poly(vinyl chloride) membrane electrodeAnalytical Chemistry, 1984
- Determination of glutamate pyruvate transaminase and pyruvate with an amperometric pyruvate oxidase sensorAnalytica Chimica Acta, 1980