The v-gel Advanced Dog is inferior to the endotracheal tube for sealing the airway in healthy canine anesthesia
Authors:
Open Access
- 1 June 2025
- journal article
- Published by American Veterinary Medical Association (AVMA) in American Journal of Veterinary Research
- Vol. 86 (6) , 1-7
Abstract
Objective: To assess the v-gel Advanced Dog supraglottic airway device (SGAD) safety and efficacy compared to a high-volume, low-pressure endotracheal tube (ETT). Methods: In a prospective randomized study, 30 client-owned mesocephalic canine patients (American Society of Anesthesiologists physical status I or II, scheduled for elective surgery) were to be enrolled and assigned to the SGAD or ETT group (15 SGAD; 15 ETT) by blocked randomization. Endotracheal tube cuffs were inflated to 25 cm H2O with a cuff inflator. The primary outcome was anesthetic circuit pressure decrease (leak) of inspiratory air at escalating anesthetic system pressures, tested at 0, 10, 15, and 20 minutes after airway device placement. Results: Due to safety concerns, the study was terminated after 6 patients in the SGAD group failed leak pressure testing under 20 cm H2O at all time points. Thirteen canines (6 SGAD, 7 ETT) were enrolled from September through October 2023. The risk of leaking for the SGAD group was significantly higher than that of the ETT group at 12 cm H2O (P = .005) and at 16 cm H2O (P = .001). There were no perioperative or postoperative adverse events or significant differences in characteristics between airway device groups except that the risk for not requiring manipulations to obtain and maintain an effective airway for the SGAD group was 67% less (relative risk, 0.33; exact 90% CI, 0.06 to 0.74) than that of the ETT group. Conclusions: The SGAD, but not ETT, failed leak pressure testing < 20 cm H2O. Clinical Relevance: The v-gel Advanced Dog SGAD was not safe to deliver inhalant anesthetics due to the device failing leak pressure testing under 20 cm H2O.Keywords
This publication has 13 references indexed in Scilit:
- Success of placement and complications during v-gel placement and maintenance of anaesthesiaJournal of Feline Medicine and Surgery, 2021
- Comparison of endoscopic endotracheal intubation and the v‐gel supraglottic airway device for spontaneously ventilating New Zealand white rabbits undergoing ovariohysterectomyVeterinary Record, 2020
- Experimental evaluation of four airway devices in anaesthetized New Zealand White rabbitsVeterinary Anaesthesia and Analgesia, 2017
- Comparison of a Supraglottic Airway Device (v-gel®) with Blind Orotracheal Intubation in RabbitsFrontiers in Veterinary Science, 2017
- Breathing systems and ancillary equipment2016
- Assessment of v‐gel supraglottic airway device placement in cats performed by inexperienced veterinary studentsVeterinary Record, 2015
- A comparison between the v-gel supraglottic airway device and the cuffed endotracheal tube for airway management in spontaneously breathing cats during isoflurane anaesthesiaVeterinary Anaesthesia and Analgesia, 2013
- Initial feasibility investigation of the v-gel airway: an anatomically designed supraglottic airway device for use in companion animal veterinary anaesthesiaVeterinary Anaesthesia and Analgesia, 2010
- G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciencesBehavior Research Methods, 2007
- Randomization and allocation concealment: a practical guide for researchersJournal of Critical Care, 2005