Abstract
Initial stability is essential for open reduction internal fixation of intraarticular calcaneal fractures. Geometrical feature of a calcaneal plate is influential to its endurance under physiological load. It is unclear if conventional and pre-contoured anatomical calcaneal plates may exhibit differently in biomechanical perspective. A Sanders’ Type II-B intraarticular calcaneal fracture model was reconstructed to evaluate the effectiveness of calcaneal plates using finite element methods. Incremental vertical joint loads up to 450 N were exerted on the subtalar joint to evaluate the stability and safety of the calcaneal plates and bony structure. Results revealed that the anatomical calcaneal plate model had greater average structural stiffness (585.7 N/mm) and lower von Mises stress on the plate (774.5 MPa) compared to those observed in the conventional calcaneal plate model (stiffness: 430.9 N/mm; stress on plate: 867.1 MPa). Although both maximal compressive and maximal tensile stress and strain were lower in the anatomical calcaneal plate group, greater loads on fixation screws were found (average 172.7 MPa compared to 82.18 MPa in the conventional calcaneal plate). It was noted that high magnitude stress concentrations would occur where the bone plate bridges the fracture line on the lateral side of the calcaneus bone. Sufficient fixation strength at the posterolateral calcaneus bone is important for maintaining subtalar joint load after reduction and fixation of a Sanders’ Type II-B calcaneal fracture. In addition, geometrical design of a calcaneal plate should worth considering for the mechanical safety in practical usage.Keywords
This publication has 17 references indexed in Scilit:
- Biomechanical analysis of foot with different foot arch heights: a finite element analysisComputer Methods in Biomechanics and Biomedical Engineering, 2012
- Biomechanical Evaluation of a Locking and Nonlocking Reconstruction Plate in an Osteoporotic Calcaneal Fracture ModelFoot & Ankle International, 2011
- Stability of Locking and Non-Locking Plates in an Osteoporotic Calcaneal Fracture ModelFoot & Ankle International, 2011
- Using an Optimization Approach to Design an Insole for Lowering Plantar Fascia Stress—A Finite Element StudyAnnals of Biomedical Engineering, 2008
- Finite element analysis of plantar fascia under stretch—The relative contribution of windlass mechanism and Achilles tendon forceJournal of Biomechanics, 2008
- Polyaxially-locked plate screws increase stability of fracture fixation in an experimental model of calcaneal fractureThe Journal of Bone and Joint Surgery. British volume, 2006
- Calcaneal FracturesThe Journal of Trauma: Injury, Infection, and Critical Care, 1996
- Operative vs. Nonoperative Treatment of Intra-Articular Fractures of the Calcaneus: A Prospective Randomized TrialFoot & Ankle International, 1996
- Extensive Intraarticular Fractures of the FootClinical Orthopaedics and Related Research, 1993
- Comparison of Open Versus Closed Reduction of Intraarticular Calcaneal FracturesJournal of Orthopaedic Trauma, 1992